Dimension Vista Protein 1 Control M, PROT 1 CON M, Siemens Material Number (SMN) 10445920, Product Code...
Siemens Healthcare Diagnostics, Inc. began a recall of Dimension Vista Protein 1 Control M, PROT 1 CON M, Siemens Material Number (SMN) 10445920, Product Code KC716; IVD on . FDA classified it as Class II and gives the reason as "Complaint of high amount of ceruloplasmin patient results below the reference range (< 20 mg/dL [0.2 g/L]. PROT1 CAL and corresponding PROT1 CON L, M, and H are inaccurately assigned for ceruloplasmin (CER)". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0175-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Dimension Vista Protein 1 Control M, PROT 1 CON M, Siemens Material Number (SMN) 10445920, Product Code KC716; IVD
- Product code
- CHN Immunochemical, Ceruloplasmin
- Reason for recall
- Complaint of high amount of ceruloplasmin patient results below the reference range (< 20 mg/dL [0.2 g/L]. PROT1 CAL and corresponding PROT1 CON L, M, and H are inaccurately assigned for ceruloplasmin (CER).
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Siemens issued an Urgent Medical Device Correction (UMDC) letter (PP-17-012.A.US, dated May 23, 2017) to US customers on May 23, 2017 and an Urgent Field Safety Notice (UFSN) to customers Outside the US in May 2017. Customers were instructed in the letter to use re-assigned CER values for the lots indicated of PROT1 CAL and PROT1 CON L, M, and H. In the US, the method of distribution is FEDEX and outside the US, the distribution is determined at the country level. Customers were requested to complete Effectiveness Check questionnaire attached to the UMDC/UFSN and return to Siemens within 30 days.
- Distribution
- Worldwide Distribution - US (nationwide) Internationally to Austria, Australia, Belgium, Canada, France, Germany, Great Britain, Italy, Japan, Netherlands, Portugal, Slovakia, Spain, Switzerland and Republic of Korea South
- Product codes and lots
- Lot numbers: 6BQM01, 6BQM01A, 6BQM01B, 6BQM01C,6BQM01D, 6BQM01E (exp. date 2017-06-17); 6FQM01, 6FQM01A (exp. date 2017-11-03) --- Lot # (UDI): 6BQM01 (008427680063336JQL01C20180131), 6BQM01A (008427680063406BQM01A20170617), 6BQM01B (008427680063406BQM01B20170617), 6BQM01C (008427680063406BQM01C20170617), 6BQM01D( 008427680063406BQM01D20170617), 6BQM01E (008427680063406BQM01E20170617), 6FQM01(008427680063406FQM0120171103), 6FQM01A (008427680063406FQM01A20171103).
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0173-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0174-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0175-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0176-2018 | Class II | Terminated | Voluntary: Firm initiated |