Jelco, Hypodermic needle Pro needles 19G X1 1/2" 800/CA Model No. EL 1915 & 19G X 2" 600/CA Model No. SB...
Smiths Medical ASD Inc. began a recall of Jelco, Hypodermic needle Pro needles 19G X1 1/2" 800/CA Model No. EL 1915 & 19G X 2" 600/CA Model No. SB 5021, Rx Only, Sterile EO on . FDA classified it as Class II and gives the reason as "Smiths Medical became aware on January 11, 2017, that syringe assemblies provided by Nipro Medical Corporation contained pin holes in the side wall of the hub. The defect was not detected during Smiths Medical incoming...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2670-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- Jelco, Hypodermic needle Pro needles 19G X1 1/2" 800/CA Model No. EL 1915 & 19G X 2" 600/CA Model No. SB 5021, Rx Only, Sterile EO.
- Product code
- FMI Needle, Hypodermic, Single Lumen
- Reason for recall
- Smiths Medical became aware on January 11, 2017, that syringe assemblies provided by Nipro Medical Corporation contained pin holes in the side wall of the hub. The defect was not detected during Smiths Medical incoming inspection. The syringe assemblies were used to manufacture seven (7) finished good lots of Hypodermic Needle Pro needles, Smiths Medical part numbers EL 1915 and SB5021.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Smiths Medical sent an "Urgent Medical Device Recall Notice" dated April 04, 2017. The letter was addressed to Quality Assurance Manager & Associate Sourcing Consultant. The letter listed the Affected devices described the Reason for Recall, & Risk to Health. Requested consignee to complete Urgent Medical Device Recall Response Form within 10 days of receipt and send it to: [email protected]. For further questions, please call (603) 352-3812
- Quantity in commerce
- 736,000
- Distribution
- US Distribution to the state of : IN
- Product codes and lots
- 3297460,3297461 Model No. EL1915. 3297458,3297459,3297465, 3297466,3297467 Model No. SB5021.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2670-2017 | Class II | Terminated | Voluntary: Firm initiated |