Z-2322-2017Class II
Philips DuraDiagnost X-ray systems: DuraDiagnost (R1.0, Product Number 712210) DuraDiagnost (R2.0, Product...
Philips North America LLC / initiated 2017-05-31 / Terminated / root cause: design
Philips Electronics North America Corporation began a recall of Philips DuraDiagnost X-ray systems: DuraDiagnost (R1.0, Product Number 712210) DuraDiagnost (R2.0, Product Number 712211) DuraDiagnost (R3.0, Product Number... on . FDA classified it as Class II and gives the reason as "Tube arm assembly could fall down due to cracked welding joints". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2322-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Philips DuraDiagnost X-ray systems: DuraDiagnost (R1.0, Product Number 712210) DuraDiagnost (R2.0, Product Number 712211) DuraDiagnost (R3.0, Product Number 712211)
- Product code
- KPR System, X-Ray, Stationary
- Reason for recall
- Tube arm assembly could fall down due to cracked welding joints
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Philips Healthcare sent an URGENT - Field Safety Notice Medical Device Correction letter dated May 23, 2017, advising users of the problem and asking facilities to stop using the device until a Philips Engineer can go on-site to perform the FC071200173; contact your local Philips representative if you have any concern or support required on this issue. If you need any further information or support concerning this issue, contact your local Philips representative. For further questions, please call (978) 659-3000.
- Quantity in commerce
- 98 sytems
- Distribution
- Worldwide Distribution - US Distribution to the states of : Ca and Oh., and to the countries of: Australia, Libya, China, Mayotte, Mexico, France, United Kingdom, Germany, Ecuador, Australia, Spain, Switzerland, Macedonia, Uzbekistan, Kuwait, Malaysia, Netherlands, Argentina, Chile, Colombia, Egypt, United Arab Emirates, Saudi Arabia, Hungary, Peru, Poland, Indonesia, Burkina Faso, Kazakhstan, Turkey and Slovenia.
- Product codes and lots
- SN120001-SN120003, SN130001-SN130023, SN140001-SN140022, SN140024, SN140027, SN140030-SN140031, SN140033-SN140046, SN140048, SN140050-SN140068, SN150001-SN150004, SN150006, SN150008-SN150011, SN160001, SN160002
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2322-2017 | Class II | Terminated | Voluntary: Firm initiated |