Z-2624-2017Class II
LEGION(R) FEM CONE IMPACTOR HEADS: a) REF 71441976, ID 18MM b) REF 71441977, ID 20MM c) REF 71441978, ID 22MM
Smith & Nephew, Inc. / initiated 2017-05-30 / Terminated / root cause: manufacturing process
Smith & Nephew, Inc. began a recall of LEGION(R) FEM CONE IMPACTOR HEADS: a) REF 71441976, ID 18MM b) REF 71441977, ID 20MM c) REF 71441978, ID 22MM on . FDA classified it as Class II and gives the reason as "The incorrect Loctite adhesive was used to assemble the impactor heads". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2624-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Smith & Nephew, Inc.
- Product
- LEGION(R) FEM CONE IMPACTOR HEADS: a) REF 71441976, ID 18MM b) REF 71441977, ID 20MM c) REF 71441978, ID 22MM
- Product code
- JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
- Reason for recall
- The incorrect Loctite adhesive was used to assemble the impactor heads.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- The firm notified the consignee by email on 05/30/2017. The notice directed the consignee to inspect its inventory and locate any unused devices from the listed product and batch numbers, and quarantine them immediately. Distributors were directed to notify their customers of the field action and ensure the recall directions were followed.
- Quantity in commerce
- 36 units
- Distribution
- US distribution to TN
- Product codes and lots
- Lot numbers: a) 16HGA0031 b) 16HGA0034 c) 16HGA0035
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2624-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2625-2017 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
