510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2659-2017Class II

3DKNEE(TM) SYSTEM e+, Tibial Insert, RIGHT Sz 8 / 11mm, REF 391-11-708, QTY 01 The 3DKnee Insert HXL Tibial...

Encore Medical, LP / initiated 2017-05-31 / Terminated / root cause: packaging

Encore Medical, Lp began a recall of 3DKNEE(TM) SYSTEM e+, Tibial Insert, RIGHT Sz 8 / 11mm, REF 391-11-708, QTY 01 The 3DKnee Insert HXL Tibial Inserts are manufactured from ultra high molecular... on . FDA classified it as Class II and gives the reason as "Mis-pack. A size 6 Left 3DKnee Tibial Insert box and pouch contained a Size 8 right 3Dknee Tibial Insert". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-2659-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Encore Medical, LP
Product
3DKNEE(TM) SYSTEM e+, Tibial Insert, RIGHT Sz 8 / 11mm, REF 391-11-708, QTY 01 The 3DKnee Insert HXL Tibial Inserts are manufactured from ultra high molecular weight polyethylene (UHMWPE) infused with pure liquid pharmaceutical grade alpha-tocopheral and then cross-linked to conform to ASTM F2695. This insert is intended to more closely match the kinematics of the knee, allowing some rotation along the medial condyle and increased congruency along the lateral condyle.
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
Mis-pack. A size 6 Left 3DKnee Tibial Insert box and pouch contained a Size 8 right 3Dknee Tibial Insert.
Root cause tag
packaging
FDA root cause
Packaging process control
Action
The firm, DJO Global, sent an "URGENT FIELD SAFETY NOTICE" dated 5/31/2017 by email to its consignees on 5/31/2017. The notice described the product, problem and actions to be taken. The consignees were instructed to do the following actions by June 16, 2017: pass notice on to all those within the organization or to any organization where the affected product has been transferred; contact customer service at 1-800-456-8696 to place a replacement order and obtain shipment information; return all affected devices; and complete, sign and return the Acknowledgment and Receipt Form by email to: [email protected]. The Regulatory Affairs department will follow-up via phone and email with non-responding consignees. If you have any questions, please contact Sr. Director Regulatory Affairs by phone at (760) 734-3551 or by email at [email protected].
Quantity in commerce
9 units
Distribution
Worldwide Distribution-US (nationwide) to states of: AL, TX, UT, and CA; and countries of: INDIA and GERMANY.
Product codes and lots
Lot # 339G1045

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2659-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2658-2017Class IITerminatedVoluntary: Firm initiated
Z-2659-2017Class IITerminatedVoluntary: Firm initiated