Z-2863-2017Class II

Spectranetics Turbo-Elite Laser Atherectomy Catheter Peripheral Rapid Exchange Catheters Device is indicated...

Spectranetics Corp. / initiated 2017-06-06 / Terminated / root cause: design

Spectranetics Corporation began a recall of Spectranetics Turbo-Elite Laser Atherectomy Catheter Peripheral Rapid Exchange Catheters Device is indicated for use in the treatment, including atherectomy... on . FDA classified it as Class II and gives the reason as "Potential for flouroscopially visible marker band at the tip of the Rapid Exchange Turbo Elite Atherectomy Catheters to become loose or disengaged rom the catheter when the device is in use in an off-label manner". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2863-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Spectranetics Corp.
Product
Spectranetics Turbo-Elite Laser Atherectomy Catheter Peripheral Rapid Exchange Catheters Device is indicated for use in the treatment, including atherectomy, of infrainguinal stenosis and occlusions
Product code
MCW Catheter, Peripheral, Atherectomy
Reason for recall
Potential for flouroscopially visible marker band at the tip of the Rapid Exchange Turbo Elite Atherectomy Catheters to become loose or disengaged rom the catheter when the device is in use in an off-label manner.
Root cause tag
design
FDA root cause
Device Design
Action
Spectranetics sent a Field Safety Notice letter via certified mail on June 7, 2017, to all customers who have ordered the affected model numbers since 23May2016 reiterating the risks associated with off-label use of the device (i.e., not activating the laser within a stent and not until all contrast media is flushed from the treatment area). The product continues to meet specifications and does not need to be returned. Customers with questions were encouraged to call their local Spectranetics Sales Representative or Spectranetics Customer Service Department. Customer Service can be reached by phone at 800-231-0978 (Option 2), by fax at 877-447-2022, or by email at [email protected]. Customer Service is available Monday thru Friday, From 7AM-5PM MST.
Distribution
Worldwide Distribution - US including AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, IL, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, Hawaii and Puerto Rico Foreign distribution to Bahrain, Brazil, Canada, Italy, Japan, and Netherlands
Product codes and lots
Model Number 414-159 Catheter Diameter 1.4mm,  Model Number 417-156 Catheter Diameter 1.7mm, and Model Number 420-159 Catheter Diameter 2.0mm.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2863-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2863-2017Class IITerminatedVoluntary: Firm initiated