Z-2693-2017Class II

Infinity 5, Implantable Pulse Generator, REF 6660, Rx only, STERILE EO

Abbott Medical / initiated 2017-06-02 / Terminated / root cause: design

St. Jude Medical, Inc. began a recall of Infinity 5, Implantable Pulse Generator, REF 6660, Rx only, STERILE EO, Product Usage: The Infinity IPG is an electronic device designed to be connected to one... on . FDA classified it as Class II and gives the reason as "The firm discovered a risk of loss of Neuromodulation therapy due to exposure of the IPG to monopolar electrosurgical devices (ESU) during medical procedures". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2693-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Infinity 5, Implantable Pulse Generator, REF 6660, Rx only, STERILE EO, Product Usage: The Infinity IPG is an electronic device designed to be connected to one or two extensions. It is powered by a hermetically sealed battery within a titanium case and uses microelectronic circuitry to generate constant-current electrical stimulation. The IPG is conductive on all sides, which allows the IPG case (also called a can) to be used as an anode for monopolar stimulation. The IPG communicates wirelessly with system programmers and controllers, and IPGs are available in small and large sizes to accommodate different power needs.
Product code
MHY Stimulator, Electrical, Implanted, For Parkinsonian Tremor
Reason for recall
The firm discovered a risk of loss of Neuromodulation therapy due to exposure of the IPG to monopolar electrosurgical devices (ESU) during medical procedures.
Root cause tag
design
FDA root cause
Device Design
Action
The notification material is in the form of a Physician Letter and Patient letter which will be delivered to physicians and patients in the US. The physician mailing will be initiated June 2, 2017 for delivery on June 7-9, 2017. The physicians were provided with patient management recommendations, Step for the patient controller updates, and Clinician's Manual Revised Warning. The Patient mailings will begin the week of June 12, 2017 to allow for clinician notification and are expected to be delivered within 3-5 business days. The patient letter informs of them of the potential for the loss of therapy and the steps to receive the software upgrade.
Quantity in commerce
443 units
Distribution
US Nationwide Distribution
Product codes and lots
AAD255.1 AAN997.1 AAP279.1 AAP328.1 AAP333.1 AAT509.1 AAT579.1 AAT582.1 AAT652.1 AAT783.1 AAT792.1 AAT795.1 AAT797.1 AAT804.1 AAT845.1 AAT902.1 AAT976.1 AAT987.1 AAT989.1 AAU003.1 AAU009.1 AAU022.1 AAU046.1 AAU078.1 AAU094.1 ABA946.1 ABM329.1 ABM630.1 ABT798.1 ABT841.1 ACE689.1 ACE771.1 ACE786.1 ACE838.1 ACE875.1 ACE900.1 ACE936.1 ACE939.1 ACE994.1 ACF065.1 ACK145.1 ACK161.1 ACK241.1 ACK424.1 ACK511.1 ACK514.1 ACK549.1 ACK559.1 ACK622.1 ACK629.1 ACK637.1 ACK699.1 ACK702.1 ACK739.1 ACK752.1 ACK803.1 ACK805.1 ACK806.1 ACK828.1 ACK855.1 ACP680.1 ACP686.1 ACP687.1 ACP700.1 ACP770.1 ACP780.1 ACP824.1 ACP833.1 ACP835.1 ACP869.1 ACP872.1 ACP874.1 ACP881.1 ACP883.1 ACT538.1 ACT548.1 ACT565.1 ACU349.1 ACU373.1 ACU381.1 ACU424.1 ADR679.1 ADR787.1 ADR794.1 ADR852.1 ADW762.1 ADW766.1 ADW771.1 AEH704.1 AEH733.1 AEH745.1 AEH748.1 AEH761.1 AEH764.1 AEH772.1 AEH826.1 AEH831.1 AEH856.1 AEH862.1 AEH865.1 AEH883.1 AEH951.1 AEH953.1 AEH955.1 AEH962.1 AEH972.1 AEJ054.1 AEJ100.1 AEJ101.1 AEJ106.1 AEJ111.1 AEJ145.1 AEJ148.1 AEJ164.1 AEJ167.1 AEJ203.1 AEJ241.1 AEJ251.1 AEJ252.1 AEJ257.1 AEM803.1 AEM833.1 AEM852.1 AEM857.1 AEM877.1 AEM885.1 AEM887.1 AEM888.1 AEM917.1 AEM951.1 AEM964.1 AEM967.1 AEM969.1 AEM988.1 AFB225.1 AFB349.1 AFB376.1 AFB425.1 AFB433.1 AFB434.1 AFB436.1 AFB444.1 AFB454.1 AFB495.1 AFB658.1 AFG344.1 AFG446.1 AFG471.1 AFG496.1 AFG545.1 AFG668.1 AFG672.1 AFG679.1 AFG704.1 AFG724.1 AFG947.1 AFG959.1 AFZ174.1 AFZ211.1 AFZ224.1 AFZ260.1 AFZ359.1 AFZ364.1 AFZ384.1 AFZ409.1 AFZ427.1 AFZ532.1 AFZ588.1 AFZ590.1 AFZ807.1 AFZ835.1 AFZ851.1 AFZ996.1 AGM602.1 AGM732.1 AGM765.1 AGM950.1 AGN231.1 AGN324.1 AGN447.1 AGN490.1 AGN552.1 AGN616.1 AHD572.1 AHD578.1 AHD589.1 AHD612.1 AHD646.1 AHD648.1 AHD663.1 AHD677.1 AHD686.1 AHD713.1 AHD818.1 AHD821.1 AHD822.1 AHK567.1 AHK727.1 AHK778.1 AHK897.1 AHK906.1 AHK922.1 AHK976.1 AHL049.1 AHL054.1 AHL069.1 AHL093.1 AHL095.1 AHL097.1 AHL105.1 AHL122.1 AHL133.1 AHL135.1 AHL138.1 AHL195.1 AHV557.1 AHV591.1 AHV603.1 AHV679.1 AHV705.1 AHV797.1 AHV801.1 AHV802.1 AHV807.1 AHV810.1 AHV818.1 AHV824.1 AHV843.1 AHV851.1 AHV854.1 AHV855.1 AHV860.1 AHV873.1 AHV921.1 AHV937.1 AHV945.1 AHV947.1 AHV965.1 AHV971.1 AHV975.1 AHV992.1 AHW028.1 AHW049.1 AHW055.1 AHW080.1 AHW082.1 AHW085.1 AHW104.1 AHW156.1 AHW489.1 AHW492.1 AHW496.1 AHW498.1 AHW661.1 AHW663.1 AHW670.1 AHW683.1 AJP994.1 AJP995.1 AJR053.1 AJR058.1 AJR073.1 AJR075.1 AJR076.1 AJR077.1 AJW280.1 AJW285.1 AJW287.1 AJW290.1 AJW296.1 AJW310.1 AJW328.1 AJW329.1 AJW334.1 AJW336.1 AJW345.1 AJW365.1 AJW382.1 AJW384.1 AJW388.1 AJW393.1 AJW412.1 AJW422.1 AJW431.1 AJW461.1 AJW487.1 AKX628.1 ALE668.1 ALE944.1 ALJ257.1 ALJ402.1 ALJ928.1 ALJ984.1 ALJ992.1 ALN367.1 ALN378.1 ALN429.1 ALN438.1 ALN441.1 ALN444.1 ALN448.1 ALN451.1 ALN469.1 ALN477.1 ALN487.1 ALN547.1 ALN593.1 ALN596.1 ALN622.1 ALN628.1 ALN634.1 ALN658.1 ALX037.1 ALX043.1 ALX053.1 ALX085.1 ALX203.1 ALX313.1 ALY252.1 ALY270.1 ALY272.1 ALY313.1 ALY329.1 ALY367.1 ALY369.1 ALY371.1 ALY400.1 ALY431.1 ALY533.1 ALY544.1 ALY610.1 ALY624.1 ALY625.1 ALY636.1 ALY653.1 ALY660.1 ALY774.1 ALY778.1 ALY799.1 ALY801.1 ALY802.1 ALY810.1 ALZ039.1 ALZ051.1 ALZ064.1 ALZ080.1 ALZ311.1 AMK053.1 AMN786.1 AMN963.1 AMP153.1 AMP226.1 AMS002.1 AMS004.1 AMS007.1 AMS022.1 AMS024.1 AMS026.1 AMS038.1 AMS068.1 AMS073.1 AMS126.1 AMS127.1 AMS161.1 AMS169.1 AMS193.1 AMS228.1 AMS241.1 AMS248.1 AMS256.1 AMS258.1 AMS259.1 AMS270.1 AMS271.1 AMS400.1 AMS431.1 AMS879.1 AMT051.1 AMT260.1 AMT351.1 ANG969.1 ANG990.1 ANH036.1 ANH126.1 ANH170.1 ANH185.1 ANH258.1 ANH278.1 ANH468.1 ANH503.1 ANH603.1 ANH618.1 ANH636.1 ANL856.1 YY2921.1 YY5780.1 YY5876.1 YY6049.1 YY6118.1 YZ1385.1 YZ1487.1 YZ1509.1 YZ1511.1 YZ1514.1 YZ1515.1 YZ1531.1 YZ1574.1 YZ1595.1 YZ9624.1 YZ9636.1 YZ9648.1 YZ9651.1 YZ9654.1 YZ9683.1 YZ9691.1 YZ9702.1 YZ9709.1 YZ9712.1 YZ9713.1 YZ9724.1 YZ9732.1 YZ9749.1 YZ9755.1 YZ9768.1 YZ9826.1 YZ9857.1 ZK3232.1 ZK3240.1 ZK3253.1 ZK3370.1 ZV3919.1 ZW0254.1 ZW0302.1 ZW0587.1 ZW8865.1 ZW8885.1 ZW8943.1 ZW8991.1 ZY8811.1 ZY8821.1 ZY9321.1 ZY9369.1 ZY9406.1

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2693-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2691-2017Class IITerminatedVoluntary: Firm initiated
Z-2692-2017Class IITerminatedVoluntary: Firm initiated
Z-2693-2017Class IITerminatedVoluntary: Firm initiated
Z-2694-2017Class IITerminatedVoluntary: Firm initiated
Z-2695-2017Class IITerminatedVoluntary: Firm initiated
Z-2696-2017Class IITerminatedVoluntary: Firm initiated