Infinity 7, Implantable Pulse Generator, REF 6663, Rx only, STERILE EO
St. Jude Medical, Inc. began a recall of Infinity 7, Implantable Pulse Generator, REF 6663, Rx only, STERILE EO Product Usage: The Infinity IPG is an electronic device designed to be connected to one... on . FDA classified it as Class II and gives the reason as "The firm discovered a risk of loss of Neuromodulation therapy due to exposure of the IPG to monopolar electrosurgical devices (ESU) during medical procedures". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2696-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- Infinity 7, Implantable Pulse Generator, REF 6663, Rx only, STERILE EO Product Usage: The Infinity IPG is an electronic device designed to be connected to one or two extensions. It is powered by a hermetically sealed battery within a titanium case and uses microelectronic circuitry to generate constant-current electrical stimulation. The IPG is conductive on all sides, which allows the IPG case (also called a can) to be used as an anode for monopolar stimulation. The IPG communicates wirelessly with system programmers and controllers, and IPGs are available in small and large sizes to accommodate different power needs.
- Product code
- MHY Stimulator, Electrical, Implanted, For Parkinsonian Tremor
- Reason for recall
- The firm discovered a risk of loss of Neuromodulation therapy due to exposure of the IPG to monopolar electrosurgical devices (ESU) during medical procedures.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The notification material is in the form of a Physician Letter and Patient letter which will be delivered to physicians and patients in the US. The physician mailing will be initiated June 2, 2017 for delivery on June 7-9, 2017. The physicians were provided with patient management recommendations, Step for the patient controller updates, and Clinician's Manual Revised Warning. The Patient mailings will begin the week of June 12, 2017 to allow for clinician notification and are expected to be delivered within 3-5 business days. The patient letter informs of them of the potential for the loss of therapy and the steps to receive the software upgrade.
- Quantity in commerce
- 40 units
- Distribution
- US Nationwide Distribution
- Product codes and lots
- AAD257.1 AAD302.1 AAP362.1 AAP375.1 AAP385.1 ALX096.1 ALX257.1 ALX258.1 ALY268.1 ALY409.1 ALY435.1 ALY632.1 ALY640.1 ALY763.1 ALY780.1 ALY788.1 ALZ034.1 ALZ042.1 ANT267.1 ANT731.1 ZV3966.1 ZW0230.1 ZW0376.1 ZW0429.1 ZW0458.1 ZW0498.1 ZW0544.1 ZW0545.1 ZW0567.1 ZW0570.1 ZW0576.1 ZW0590.1 ZX5511.1 ZY9379.1 ZY9394.1 ZY9398.1 ZY9405.1 ZY9413.1 ZY9433.1 ZY9530.1
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2691-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2692-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2693-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2694-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2695-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2696-2017 | Class II | Terminated | Voluntary: Firm initiated |