BD SafetyGlide" Injection Needle Catalog #305917 The SafetyGlide " needle is intended to be used for general...
Becton Dickinson & Company began a recall of BD SafetyGlide" Injection Needle Catalog #305917 The SafetyGlide " needle is intended to be used for general purpose injection, aspiration of fluids from parts... on . FDA classified it as Class II and gives the reason as "BD identified that several lots did not receive the specified dose of irradiation necessary to meet the Sterility Assurance Level (SAL) required". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2748-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Becton Dickinson & Co
- Product
- BD SafetyGlide" Injection Needle Catalog #305917 The SafetyGlide " needle is intended to be used for general purpose injection, aspiration of fluids from parts of the body below the surface of the skin, injection into intravenous ports and tubings external to the patient. The needlestick prevention feature of the device contains a mechanism that covers the needlepoint after use. In the activated position the needle cover guards against accidental needle sticks during normal handling and disposal of the used needle/syringe combination
- Product code
- FMI Needle, Hypodermic, Single Lumen
- Reason for recall
- BD identified that several lots did not receive the specified dose of irradiation necessary to meet the Sterility Assurance Level (SAL) required.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- BD sent recall letter/return response form on 9/16/2016.
- Quantity in commerce
- 29,500 units
- Distribution
- Nationwide Distribution
- Product codes and lots
- 6022838 2021-01
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2747-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2748-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2749-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2750-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2751-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2752-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2753-2017 | Class II | Terminated | Voluntary: Firm initiated |