Z-0191-2018Class II

SIEMENS N/T Protein Control SL for use on the BN II, BN ProSpec Systems: N/T Protein Control SL/L...

CSL Behring LLC / initiated 2017-06-05 / Terminated / root cause: manufacturing process

CSL Behring GmbH began a recall of SIEMENS N/T Protein Control SL for use on the BN II, BN ProSpec Systems: N/T Protein Control SL/L: REF/Product Code OQIN19, SMN 10446079 (Low); N/T Protein... on . FDA classified it as Class II and gives the reason as "Siemens Healthcare Diagnostics has confirmed a negative bias of approximately -20% for results obtained with N Antiserum to Human Ceruloplasmin when calibrating Ceruloplasmin with various lots of N Protein Standard SL...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0191-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
CSL Behring LLC
Product
SIEMENS N/T Protein Control SL for use on the BN II, BN ProSpec Systems: N/T Protein Control SL/L: REF/Product Code OQIN19, SMN 10446079 (Low); N/T Protein Control SL/M, REF/Product Code OQIO19, SMN 10446085 (Mid); N/T Protein Control SL/H, REF/Product Code OQIP19, SMN 10446089 (High);
Product code
CHN Immunochemical, Ceruloplasmin
Reason for recall
Siemens Healthcare Diagnostics has confirmed a negative bias of approximately -20% for results obtained with N Antiserum to Human Ceruloplasmin when calibrating Ceruloplasmin with various lots of N Protein Standard SL. N/T Protein Controls SL for Ceruloplasmin are similarly biased. --- The affected standards and controls for the affected lots were incorrectly value assigned, which led to a lower result compared to correctly assigned standards. Note: N Antiserum to Human Ceruloplasmin (REF OUIE) works as intended. --- Siemens investigation indicated that N Protein Standard SL and corresponding N/T Protein Control SL (Levels Low, Mid and High) are inaccurately assigned for Ceruloplasmin.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
URGENT MEDICAL DEVICE CORRECTION Letter (PP17-011.a.US, dated May 30, 2017) was sent to US consignees via FedEx on 6/5/2017. URGENT FIELD CORRECTIVE ACTION Letter was e-mailed to customers Outside the US. Customers are instructed to use the re-assigned Ceruloplasmin values for the lots indicated in the letter.
Distribution
Worldwide including the countries of United States, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia-Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Costa Rica, Croatia, Czech Republic, Cyprus, Denmark, Egypt, Finland, France, Germany, Greece, Great Britain, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, japan, Kazakhstan, Kuwait, Latvia, Lebanon, Luxembourg, Macedonia, Mexico, Morocco, Netherlands, New Zealand, Norway, Poland, Portugal, Republic of Korea (S), Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Spain, Sweden, South Africa, Switzerland, Syria, Taiwan, Thailand, Trinidad-Tobago, Turkey, Turkmenistan, United Arab Emirates, Uruguay, Vatican, Venezuela,
Product codes and lots
SMN 10446079: Lot (UDI): 084647A (00842768016127084647A20170913), 084647C (00842768016127084647C20170913) exp. date 2017-09-13; 084648(0084276801612708464820180430),  084648C (00842768016127084648C20180430),  084648H (00842768016127084648H20180430), exp. date 2018-04-30. --- SMN 10446085: Lot (UDI): 084749D (00842768016134084749D20170917) exp. date 2017-09-17; 084750 (03 0084276801613408475020180203),  084750B (00842768016134084750B20180203), 084750E (00842768016134084750E20180203) exp. date 2018-02-03. --- SMN 10446089: Lot (UDI) 084847A (00842768016110084847A20170613),  084847B (00842768016110084847B20170613) exp. date 2017-06-13; 084848A (00842768016110084848A20171217),  084848D (00842768016110084848D20171217),  084848E (00842768016110084848E20171217),  084848H (00842768016110084848H20171217),  084848L (00842768016110084848L20171217) exp. date 2017-12-17;  084849 (0084276801611008484920180704),  084849C (00842768016110084849C20180704), 084849D (00842768016110084849D20180704) 084849E (00842768016110084849E20180704) exp. date 2018-07-04.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0191-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0190-2018Class IITerminatedVoluntary: Firm initiated
Z-0191-2018Class IITerminatedVoluntary: Firm initiated