Z-2736-2017Class I

CS 100i Intra-Aortic Balloon Pump

Datascope Corp. / initiated 2017-06-16 / Terminated / root cause: design

Maquet Datascope Corp - Cardiac Assist Division began a recall of CS 100i Intra-Aortic Balloon Pump on . FDA classified it as Class I and gives the reason as "The device failed to pump due to an electrical test failure code #58 (power up vent test fail), maintenance code #3, and an autofill failure which has been associated to a patient death due to the failure of the device...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2736-2017
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Datascope Corp.
Product
CS 100i Intra-Aortic Balloon Pump
Product code
DSP System, Balloon, Intra-Aortic And Control
Reason for recall
The device failed to pump due to an electrical test failure code #58 (power up vent test fail), maintenance code #3, and an autofill failure which has been associated to a patient death due to the failure of the device to initiate therapy.
Root cause tag
design
FDA root cause
Component design/selection
Action
Datascope Corp/MAQUET issued a press release on June 16, 2017. Datascope Corp/MAQUET sent an Urgent Product Recall Letter on June 19, 2017, to their customers via FedEx next day delivery service. Pursuant to the WARNINGS section of our CS100i, CS100 or CS300 IABP Operating/User Instructions, clinicians are instructed not to leave the patient unattended during IABP therapy. Until the service is performed, we recommend the IABP be powered on prior to initiation of therapy to allow the IABP to successfully complete its self-test. This action will take less than 60 seconds to perform. In the event the IABP fails to successfully complete the self-test and exhibits electrical test failure code 58, please remove the IABP from service and contact your local Maquet/Getinge Sales & Service Office. Customers will be instructed to complete and return the attached Medical Device Field Correction Response Form via fax to 1-973-807-9217 or send via email to [email protected]@getinge.com. If you are a distributor who has shipped any affected products to customers, please forward this document to their attention for appropriate action. For U.S. customers with technical questions, please contact our Technical Support Department (at 1-888-627-8383 and press 3), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. EST
Quantity in commerce
12,319 units total
Distribution
Worldwide Distribution - (in over 100 countries) US (Nationwide)
Product codes and lots
0998-UC-0446HXX 0998-UC-0479HXX

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2736-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2736-2017Class ITerminatedVoluntary: Firm initiated
Z-2737-2017Class ITerminatedVoluntary: Firm initiated
Z-2738-2017Class ITerminatedVoluntary: Firm initiated