CS 100i Intra-Aortic Balloon Pump
Maquet Datascope Corp - Cardiac Assist Division began a recall of CS 100i Intra-Aortic Balloon Pump on . FDA classified it as Class I and gives the reason as "The device failed to pump due to an electrical test failure code #58 (power up vent test fail), maintenance code #3, and an autofill failure which has been associated to a patient death due to the failure of the device...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2736-2017
- Classification
- Class I
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Datascope Corp.
- Product
- CS 100i Intra-Aortic Balloon Pump
- Product code
- DSP System, Balloon, Intra-Aortic And Control
- Reason for recall
- The device failed to pump due to an electrical test failure code #58 (power up vent test fail), maintenance code #3, and an autofill failure which has been associated to a patient death due to the failure of the device to initiate therapy.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- Datascope Corp/MAQUET issued a press release on June 16, 2017. Datascope Corp/MAQUET sent an Urgent Product Recall Letter on June 19, 2017, to their customers via FedEx next day delivery service. Pursuant to the WARNINGS section of our CS100i, CS100 or CS300 IABP Operating/User Instructions, clinicians are instructed not to leave the patient unattended during IABP therapy. Until the service is performed, we recommend the IABP be powered on prior to initiation of therapy to allow the IABP to successfully complete its self-test. This action will take less than 60 seconds to perform. In the event the IABP fails to successfully complete the self-test and exhibits electrical test failure code 58, please remove the IABP from service and contact your local Maquet/Getinge Sales & Service Office. Customers will be instructed to complete and return the attached Medical Device Field Correction Response Form via fax to 1-973-807-9217 or send via email to [email protected]@getinge.com. If you are a distributor who has shipped any affected products to customers, please forward this document to their attention for appropriate action. For U.S. customers with technical questions, please contact our Technical Support Department (at 1-888-627-8383 and press 3), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. EST
- Quantity in commerce
- 12,319 units total
- Distribution
- Worldwide Distribution - (in over 100 countries) US (Nationwide)
- Product codes and lots
- 0998-UC-0446HXX 0998-UC-0479HXX
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2736-2017 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-2737-2017 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-2738-2017 | Class I | Terminated | Voluntary: Firm initiated |