510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0096-2018Class II

ROSA Surgical Device 2.5.8 ROSA Surgical Device is a computer-controlled electromechanical arm. It is...

Zimmer GmbH / initiated 2013-12-18 / Terminated / root cause: design

Zimmer Biomet, Inc. began a recall of ROSA Surgical Device 2.5.8 ROSA Surgical Device is a computer-controlled electromechanical arm. It is intended to be used in the operating room for the spatial... on . FDA classified it as Class II and gives the reason as "Possible break in the connector of the Force Sensor". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0096-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
ROSA Surgical Device 2.5.8 ROSA Surgical Device is a computer-controlled electromechanical arm. It is intended to be used in the operating room for the spatial positioning and orientation of an instrument holder or tool guide. Guidance is based on a pre-operative plan developed with three-dimensional imaging software, and uses fiducial markers or optical registration. The system is intended for use by neurosurgeons to guide standard neurosurgical instruments. It is indicated for any neurosurgical condition in which the use of stereotactic surgery may be appropriate
Product code
HAW Neurological Stereotaxic Instrument
Reason for recall
Possible break in the connector of the Force Sensor.
Root cause tag
design
FDA root cause
Device Design
Action
Customers were informed onsite by Field Service Technicians of the planned correction.
Quantity in commerce
21 units total
Distribution
Nationwide Distribution to OH, MI, and TX
Product codes and lots
Model Number: ROSA 2.5.8

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0096-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0096-2018Class IITerminatedVoluntary: Firm initiated