Z-0273-2018Class II

VITROS Chemistry Products PHYT Slides, REF/CATALOG 829 8671, Rx ONLY, IVD; Ortho-Clinical Diagnostics, Inc...

Ortho-Clinical Diagnostics, Inc. / initiated 2017-06-12 / Terminated / root cause: design

Ortho-Clinical Diagnostics began a recall of VITROS Chemistry Products PHYT Slides, REF/CATALOG 829 8671, Rx ONLY, IVD; Ortho-Clinical Diagnostics, Inc. For in vitro diagnostic use only. VITROS Chemistry... on . FDA classified it as Class II and gives the reason as "The firm identified the potential for biased results to be generated when processing VITROS PHYT Slide product with citrate plasma specimen samples. The current Instructions for Use (IFU) and Assay Summary Chart state...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0273-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Product
VITROS Chemistry Products PHYT Slides, REF/CATALOG 829 8671, Rx ONLY, IVD; Ortho-Clinical Diagnostics, Inc. For in vitro diagnostic use only. VITROS Chemistry Products PHYT Slides quantitatively measure phenytoin (PHYT) concentration in serum and plasma using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.
Product code
DIP Enzyme Immunoassay, Diphenylhydantoin
Reason for recall
The firm identified the potential for biased results to be generated when processing VITROS PHYT Slide product with citrate plasma specimen samples. The current Instructions for Use (IFU) and Assay Summary Chart state serum, heparin plasma, and citrate plasma as acceptable specimen types for use with the PHYT assay.
Root cause tag
design
FDA root cause
Device Design
Action
Ortho-Clinical Diagnostic sent an Urgent Product Correction Letter dated June 12, 2017, sent via FedEx overnight courier and/ or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail (for PO Boxes only) to all customers who have a VITROS analyzer that can process PHYT Slides to inform them that negatively biased results may occur when processing citrate plasma samples using VITROS Chemistry Products PHYT Slides. Customers were instructed to immediately discontinue using citrate plasma samples to process VITROS PHYT Slides. They were also informed that it is still acceptable to use serum and heparin plasma samples to process VITROS PHYT Slides. Foreign affiliates were informed by e-mail on June 12, 2017 of the issue and instructed to notify their consignees of the issue and the appropriate actions to take. For further questions, please call (908) 218-8776.
Quantity in commerce
US; 4680; Foreign: 1286
Distribution
Worldwide Distribution - US Distribution including Puerto Rico and the following countries: Bermuda, Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Republic of Panama, Poland, Singapore, Spain, Russia and Venezuela.
Product codes and lots
Unique Identifier Number ( UDI) 10758750004690; This issue affects expired, current (in date) and future product GENs.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0273-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0273-2018Class IITerminatedVoluntary: Firm initiated