Z-2672-2017Class II
Proteus 235 and Proteus ONE proton therapy systems
Ion Beam Applications S.A. / initiated 2016-05-18 / Terminated / root cause: manufacturing process
Ion Beam Applications S.A. began a recall of Proteus 235 and Proteus ONE proton therapy systems on . FDA classified it as Class II and gives the reason as "Recall is being initiated to address on-site software installations not being performed consistently with IBA standard operating procedures". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2672-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ion Beam Applications S.A.
- Product
- Proteus 235 and Proteus ONE proton therapy systems
- Product code
- LHN System, Radiation Therapy, Charged-Particle, Medical
- Reason for recall
- Recall is being initiated to address on-site software installations not being performed consistently with IBA standard operating procedures.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- An Urgent Field Safety Notice MID-54097 was mailed to affected customers to inform that IBA plans to deploy a new software by March 2017 to correct the issue. Customers were reminded to follow standard protocol until the software updates are deployed.
- Quantity in commerce
- 18 worldwide and 10, in the U.S.
- Distribution
- Products distributed to FL, VA, PA, OK, IL, NJ, WA, TN, LA, TX, South Korea, Germany, France, Czech Republic, Italy, Poland, and Sweden
- Product codes and lots
- Code/Serial Numbers: PAT.003, PAT.006, PAT.107, PAT.108, PAT.109, PAT.110, PAT.111, PAT.112, PAT.113, PAT.114, PAT.115, SAT.116, SAT.117, SAT.119, SAT.120, SAT.122, SAT.123 and SAT.126
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2672-2017 | Class II | Terminated | Voluntary: Firm initiated |