Z-2755-2017Class II

Medtronic Visualase(R) Cooled Laser Applicator System (VCLAS) 15MM TIP, REF 9735561, (1) Laser Diffusing...

Medtronic Navigation, Inc / initiated 2017-06-07 / Terminated / root cause: design

Medtronic Navigation, Inc. began a recall of Medtronic Visualase(R) Cooled Laser Applicator System (VCLAS) 15MM TIP, REF 9735561, (1) Laser Diffusing Fiber - 600um Core, 15mm Tip, (1) Cooling Catheter... on . FDA classified it as Class II and gives the reason as "Medtronic has become aware that the VCLAS 15MM TIP cap, also referred to as Tuohy Borst Adapter (TBA), may not be capable of creating a seal around the optical fiber to ensure adequate saline flow through the cooling...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2755-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Navigation, Inc
Product
Medtronic Visualase(R) Cooled Laser Applicator System (VCLAS) 15MM TIP, REF 9735561, (1) Laser Diffusing Fiber - 600um Core, 15mm Tip, (1) Cooling Catheter with 3M(TM) SteriStrip(TM), (1) Pump Tubing Set, (1) Drainage Bag. Medtronic Navigation, Inc. Louisville, CO 80027
Product code
GEX Powered Laser Surgical Instrument
Reason for recall
Medtronic has become aware that the VCLAS 15MM TIP cap, also referred to as Tuohy Borst Adapter (TBA), may not be capable of creating a seal around the optical fiber to ensure adequate saline flow through the cooling system. When the TBA is unable to seal around the optical fiber, low retention force to the optical fiber may occur, causing saline to leak out of the TBA.
Root cause tag
design
FDA root cause
Component design/selection
Action
Medtronic sent an Urgent Medical Device Safety Alert dated June 1, 2017, to all affected customers. Actions to be taken: 1. Please examine your inventory and if any of the affected products listed above are found, immediately quarantine them for return to Medtronic. 2. Complete and follow instruction on the attached consignee response form. Sign and date the bottom of the form and then return the form to Medtronic at RS.NavFCAmedtronic.com or fax to Medtronic Technical Services at 651-367-7075. If you have affected products, contact Medtronic Technical Services at 1-800 595-9709 to receive a return material authorization (RMA) and arrange for their return and no charge replacements. Once an RMA number is obtained, ship the affected product to Medtronic. For further questions please call (720) 890-3241.
Quantity in commerce
214 units
Distribution
US Distribution to the states of : CA, GA, MN, FL, NY, IL, TX
Product codes and lots
Lot Numbers: 211158059, 211560973, 211959879, 12473141, 212707035, 211158151, 211594328, 211977586, 212503634, 212745660, 211158152, 211599898, 211987873, 212504463, 212796559, 211181212, 211628257, 212127654, 212549456, 212912501, 211181547,211908695, 212141557, 212553428, 211184941, 211916062, 212148368, 212632884,211555011, 211926889, 212434500, 212647830, 211555017, 211950736, 212438405, 212701679

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2755-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2755-2017Class IITerminatedVoluntary: Firm initiated