ROSA Surgical Device 2.5.8, a computer-controlled electromechanical arm to guide standard neurosurgical...
Zimmer Biomet, Inc. began a recall of ROSA Surgical Device 2.5.8, a computer-controlled electromechanical arm to guide standard neurosurgical instruments on . FDA classified it as Class II and gives the reason as "Potential failure of a force feedback sensor that may result in continued movement of an operating arm (rather than an immediate stop as intended)". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0160-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- ROSA Surgical Device 2.5.8, a computer-controlled electromechanical arm to guide standard neurosurgical instruments.
- Product code
- HAW Neurological Stereotaxic Instrument
- Reason for recall
- Potential failure of a force feedback sensor that may result in continued movement of an operating arm (rather than an immediate stop as intended).
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- All 9 domestic units were corrected by the firm in February 2017, via technician visits to consignee sites.
- Quantity in commerce
- 31 units
- Distribution
- US distribution in AR, GA, OH, MI, and TX.
- Product codes and lots
- Model No. 2.5.8., Serial No. RO10009, RO10011, RO10014, RO13023, RO13027, RO14031, RO14033, RO14035, and RO14037.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0160-2018 | Class II | Terminated | Voluntary: Firm initiated |