510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0160-2018Class II

ROSA Surgical Device 2.5.8, a computer-controlled electromechanical arm to guide standard neurosurgical...

Zimmer GmbH / initiated 2014-11-06 / Terminated / root cause: design

Zimmer Biomet, Inc. began a recall of ROSA Surgical Device 2.5.8, a computer-controlled electromechanical arm to guide standard neurosurgical instruments on . FDA classified it as Class II and gives the reason as "Potential failure of a force feedback sensor that may result in continued movement of an operating arm (rather than an immediate stop as intended)". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0160-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
ROSA Surgical Device 2.5.8, a computer-controlled electromechanical arm to guide standard neurosurgical instruments.
Product code
HAW Neurological Stereotaxic Instrument
Reason for recall
Potential failure of a force feedback sensor that may result in continued movement of an operating arm (rather than an immediate stop as intended).
Root cause tag
design
FDA root cause
Device Design
Action
All 9 domestic units were corrected by the firm in February 2017, via technician visits to consignee sites.
Quantity in commerce
31 units
Distribution
US distribution in AR, GA, OH, MI, and TX.
Product codes and lots
Model No. 2.5.8., Serial No. RO10009, RO10011, RO10014, RO13023, RO13027, RO14031, RO14033, RO14035, and RO14037.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0160-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0160-2018Class IITerminatedVoluntary: Firm initiated