Z-3034-2017Class II
ThermaSure(TM), a CENORIN(TM) product, SERIES 1000 MEDICAL DEVICE DRYER: (a) Model 1009; (b) Model 1010...
Cenorin, LLC / initiated 2017-06-07 / Terminated / root cause: manufacturing process
Cenorin, LLC began a recall of ThermaSure(TM), a CENORIN(TM) product, SERIES 1000 MEDICAL DEVICE DRYER: (a) Model 1009; (b) Model 1010 Forced air dryer for medical devices on . FDA classified it as Class II and gives the reason as "Dryer unit not heating. Manufacturing issue causes the terminal block to overheat, char and/or burn". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-3034-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Cenorin, LLC
- Product
- ThermaSure(TM), a CENORIN(TM) product, SERIES 1000 MEDICAL DEVICE DRYER: (a) Model 1009; (b) Model 1010 Forced air dryer for medical devices
- Product code
- JRJ Unit, Drying
- Reason for recall
- Dryer unit not heating. Manufacturing issue causes the terminal block to overheat, char and/or burn.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Cenorin initially contacted all domestic customers at the facility and distributor level via phone on June 7, 2017, which was followed by an Urgent Medical Device Recall letter dated July 20, 2017, to all affected customers. Cenorin sent initial notification to international distributors via email. Customers were asked to immediately examine their inventory for the affected product. If the product exhibits any symptoms of failure, immediately cease using it and contact Cenorin immediately at [email protected] or call Technical Support at 800-426-1042, ext 224.
- Quantity in commerce
- 18 units
- Distribution
- Worldwide Distribution - US (Nationwide) Internationally to Japan, Hong Kong, Brunei, and Israel
- Product codes and lots
- (a) Serial Numbers: 191601, 191602, 191604, 191605, 19161, 191611, 191612, 191613, 191614, 191615, 191616 (b) Serial numbers: 101505, 101506, 101603, 101606, 101607, 101608, 101609
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-3032-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-3033-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-3034-2017 | Class II | Terminated | Voluntary: Firm initiated |