Glidesheath Slender Hydrophilic Coated Introducer Sheath A-Kit, 6FR - 16CM (0.021), REF/Product Code...
Terumo Medical Corp began a recall of Glidesheath Slender Hydrophilic Coated Introducer Sheath A-Kit, 6FR - 16CM (0.021), REF/Product Code RM*ES6F16HAU, STERILE, Rx ONLY on . FDA classified it as Class II and gives the reason as "An incomplete seal at the chevron end of the primary packaging of the GLIDESHEATH SLENDER Introducer Sheath was detected in a small quantity of the affected lot population, 0.01% (1 in 10, 000 units). This breach of the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0221-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Terumo Medical Corp.
- Product
- Glidesheath Slender Hydrophilic Coated Introducer Sheath A-Kit, 6FR - 16CM (0.021), REF/Product Code RM*ES6F16HAU, STERILE, Rx ONLY
- Product code
- DYB Introducer, Catheter
- Reason for recall
- An incomplete seal at the chevron end of the primary packaging of the GLIDESHEATH SLENDER Introducer Sheath was detected in a small quantity of the affected lot population, 0.01% (1 in 10, 000 units). This breach of the sterile barrier may compromise the sterility of the sheath.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- URGENT VOLUNTARY MEDICAL DEVICE RECALL Letters (dated June 7, 2017) and Response Forms were sent to the consignees via UPS on June 7, 2017.
- Distribution
- Nationwide and Canada
- Product codes and lots
- Lot s 161209, 170118, 170215, 170317
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0214-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0215-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0216-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0217-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0218-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0219-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0220-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0221-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0222-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0223-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0224-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0225-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0226-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0227-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0228-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0229-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0230-2018 | Class II | Terminated | Voluntary: Firm initiated |