Z-0100-2018Class II
ROSA Surgical Device 2.5.8 ROSA Surgical Device is a computer-controlled electromechanical arm. It is...
Zimmer GmbH / initiated 2012-04-19 / Terminated / root cause: design
Zimmer Biomet, Inc. began a recall of ROSA Surgical Device 2.5.8 ROSA Surgical Device is a computer-controlled electromechanical arm. It is intended to be used in the operating room for the spatial... on . FDA classified it as Class II and gives the reason as "Potential failure of the optical distance sensor due to cable disconnection". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0100-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- ROSA Surgical Device 2.5.8 ROSA Surgical Device is a computer-controlled electromechanical arm. It is intended to be used in the operating room for the spatial positioning and orientation of an instrument holder or tool guide. Guidance is based on a pre-operative plan developed with three-dimensional imaging software, and uses fiducial markers or optical registration. The system is intended for use by neurosurgeons to guide standard neurosurgical instruments. It is indicated for any neurosurgical condition in which the use of stereotactic surgery may be appropriate
- Product code
- HAW Neurological Stereotaxic Instrument
- Reason for recall
- Potential failure of the optical distance sensor due to cable disconnection.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Zimmer Biomet sent an Urgent Medical Device Recall Correction letter dated July 26, 2017, to all affected customers. Customers were informed on site of the device correction by Field Service Technicians. For questions regarding this recall call 574-267-6131.
- Quantity in commerce
- 14 units
- Distribution
- Worldwide Distribution - US including OH and MI Internationally to Canada, China, France, Germany, and Italy
- Product codes and lots
- Serial No. RO10011 and RO10014
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0100-2018 | Class II | Terminated | Voluntary: Firm initiated |