ROSA Surgical Device 2.5.8

Zimmer GmbH / initiated 2013-02-10 / Terminated / root cause: software

Zimmer Biomet, Inc. began a recall of ROSA Surgical Device 2.5.8 on . FDA classified it as Class II and gives the reason as "Issue with optional neuro-endoscopy module detected under specific conditions during internal testing". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0178-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
ROSA Surgical Device 2.5.8
Product code
HAW Neurological Stereotaxic Instrument
Reason for recall
Issue with optional neuro-endoscopy module detected under specific conditions during internal testing.
Root cause tag
software
FDA root cause
Software design
Action
Field Service Technicians were dispatched to correct units by upgrading the system to ROSA 2.5.7. A further update of ROSA 2.5.8.4. A Field Safety Notice dated 07/10/2013 notifying consignees of a risk in the ROSA system recognized in internal testing. It was suggested that consignees not utilize the neuro-endoscopy module until further notice. The recall notice should have been shared with the appropriate parties and the signed acknowledgement form returned to the recalling firm.
Quantity in commerce
22 units (4 US and 18 OUS)
Distribution
AR, OH, MI, and TX
Product codes and lots
Serial No. RO10011, RO10014, RO13023 and RO13027

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0178-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0178-2018Class IITerminatedVoluntary: Firm initiated