Z-0178-2018Class II
ROSA Surgical Device 2.5.8
Zimmer GmbH / initiated 2013-02-10 / Terminated / root cause: software
Zimmer Biomet, Inc. began a recall of ROSA Surgical Device 2.5.8 on . FDA classified it as Class II and gives the reason as "Issue with optional neuro-endoscopy module detected under specific conditions during internal testing". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0178-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- ROSA Surgical Device 2.5.8
- Product code
- HAW Neurological Stereotaxic Instrument
- Reason for recall
- Issue with optional neuro-endoscopy module detected under specific conditions during internal testing.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Field Service Technicians were dispatched to correct units by upgrading the system to ROSA 2.5.7. A further update of ROSA 2.5.8.4. A Field Safety Notice dated 07/10/2013 notifying consignees of a risk in the ROSA system recognized in internal testing. It was suggested that consignees not utilize the neuro-endoscopy module until further notice. The recall notice should have been shared with the appropriate parties and the signed acknowledgement form returned to the recalling firm.
- Quantity in commerce
- 22 units (4 US and 18 OUS)
- Distribution
- AR, OH, MI, and TX
- Product codes and lots
- Serial No. RO10011, RO10014, RO13023 and RO13027
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0178-2018 | Class II | Terminated | Voluntary: Firm initiated |