Converters(R) Under Buttocks Drape with Fluid Control Pouch II, REF 8482, STERILE R, Qty. 1, Cardinal Health...
Cardinal Health 200, LLC began a recall of Converters(R) Under Buttocks Drape with Fluid Control Pouch II, REF 8482, STERILE R, Qty. 1, Cardinal Health, Waukegan, IL A surgical drape and drape... on . FDA classified it as Class II and gives the reason as "Outer label of some units of product number/lot number 8482, Lot 16LFS235 and product number/lot number 9414, Lot 16LUS317 do not match the drape packaged inside". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2757-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- Converters(R) Under Buttocks Drape with Fluid Control Pouch II, REF 8482, STERILE R, Qty. 1, Cardinal Health, Waukegan, IL A surgical drape and drape accessories is a device made of natural or synthetic materials intended to be used as a protective patient covering, such as to isolate a site of surgical incision from microbial and other contamination.
- Product code
- KKX Drape, Surgical, Antimicrobial
- Reason for recall
- Outer label of some units of product number/lot number 8482, Lot 16LFS235 and product number/lot number 9414, Lot 16LUS317 do not match the drape packaged inside.
- Root cause tag
- labeling and UDI
- FDA root cause
- Mixed-up of materials/components
- Action
- Cardinal Health sent an Urgent Product Recall letter dated June 27, 2017, to all affected consignees. In the letter, the firm requested the following actions be taken: 1. Inspect your inventory for the affected product code and lot number (listed on the package) 2. segregate and quarantine all on-hand product 3. Return the enclosed acknowledgment form via fax to 847-689-9101 ir614-652-9648, whether or not you have affected product, indicating the product code, lot and quantity of product you've quarantined or discarded. 4. Notify any customers to whom may have distributed or forwarded product affected by this product. 5. Contact the appropriate Customer Service group to arrange for return and credit of any affected product: Hospital - 800-964-5227 Federal Government - 800-444-1166 Distributor - 600-635-6021 All other customers - 888-444-5401 For questions regarding this recall call 800-327-7503.
- Quantity in commerce
- 1591 each
- Distribution
- Worldwide Distribution - US (Nationwide including PR) and Canada
- Product codes and lots
- Lot 16LFS235, expires November 1, 2021
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2757-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2758-2017 | Class II | Terminated | Voluntary: Firm initiated |