Z-2772-2017Class II
STOCKERT Heater-Cooler System 3T, 240V/60 Hz
LivaNova USA, Inc. / initiated 2017-06-30 / Terminated / root cause: design
Sorin Group USA, Inc. began a recall of STOCKERT Heater-Cooler System 3T, 240V/60 Hz on . FDA classified it as Class II and gives the reason as "Plastic heat exchanger fibers separating the blood compartment from the water compartment in certain oxygenators may allow diffusion of hydrogen peroxide in a quantity that exceeds allowable limits". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2772-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- LivaNova USA, Inc.
- Product
- STOCKERT Heater-Cooler System 3T, 240V/60 Hz
- Product code
- DWC Controller, Temperature, Cardiopulmonary Bypass
- Reason for recall
- Plastic heat exchanger fibers separating the blood compartment from the water compartment in certain oxygenators may allow diffusion of hydrogen peroxide in a quantity that exceeds allowable limits.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- "Consignees were sent a recall notification, dated 06/30/2017, via mail. Instructions included a recommendation from LivaNova to use its oxygenators, or other oxygenators that are equipped with compatible heat exchangers, to ensure that the permeability rate is within the prescribed allowable limits (ALs). Depending on whether or not the oxygenator was from LivaNova, there were specific instructions to follow for each type and source. Consignees were also instructed to complete and return the Confirmation Form. If product was further distributed, the firm asks that further customers be notified of the recall and that customers contact the LivaNova Quality Assurance Team at [email protected]. Questions regarding the recall can be directed to 800-986-4702 or [email protected]. Any adverse reactions are to be reported to FDA's MedWatch Adverse Event Reporting program."
- Quantity in commerce
- 7,546 units
- Distribution
- "Domestic: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, & WY. Foreign: AL, AE, AR, AT, AU, AZ, BD, BE, BG, BH, BR, BY, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, DZ, EC, EE, EG, ES, ET, FI, FR, GB, GE, GR, GY, HK, HR, HU, IE, ID, IL, IN, IQ, IR, IT, JO, JP, KR, KW, KZ, LB, LK, LT, LU, LV, LY, MA, MC, MK, MN, MT, MU, MX, MY, NG, NL, NO, NP, NZ, OM, PA, PE, PH, PK, PL, PR, PS, PT, QA, RE, RO, RS, RU, SA, SE, SG, SK, SV, SY, TH, TN, TR, TT, TW, UA, VN, & ZA."
- Product codes and lots
- Item No. 16-02-81; Serial No. 16S10743-16S11708
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2771-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2772-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2773-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2774-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2775-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2776-2017 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.