Medtronic Curved Suction 70 9733450 EM ENT, reusable, Rx. The firm name on the label is Medtronic Navigation...
Medtronic Navigation, Inc. began a recall of Medtronic Curved Suction 70 9733450 EM ENT, reusable, Rx. The firm name on the label is Medtronic Navigation, Inc., Louisville, CO. The suction devices are... on . FDA classified it as Class II and gives the reason as "The instruments are not able to pass instrument verification which is performed prior to the use of each instrument". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-3052-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic Navigation, Inc
- Product
- Medtronic Curved Suction 70 9733450 EM ENT, reusable, Rx. The firm name on the label is Medtronic Navigation, Inc., Louisville, CO. The suction devices are part of a family of reusable ENT instruments that are used with the StealthStationTM AxiEM surgical navigation system. The AxiEM system is an electromagnetic (EM) tracking solution for surgical tools. The system measures an induced voltage created by local magnetic fields from a set of coils in the transmit coil array (TCA). Each surgical tool has a set of receiving coils to sense the magnetic field strength. Using a mathematical algorithm, the position and orientation of the instrument can be computed and overlaid/registered to imagery on a computer monitor, resulting in the ability to track the instrument tip in relation to the patients anatomy. AxiEM ENT Suction devices are used as suction and pointing devices. The suction devices are intended to have an ENT Instrument Tracker mounted on it
- Product code
- PGW Ear, Nose, And Throat Stereotaxic Instrument
- Reason for recall
- The instruments are not able to pass instrument verification which is performed prior to the use of each instrument.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Medtronic sent an Urgent Medical Device Recall letter dated June 25, 2017, via overnight mail on June 29, 2017, to all affected customers. Letters to foreign customers are being handled by geography-specific Regulatory Specialists. Customers were instructed examine their inventory for the affected product and immediately quarantine them for return for Medtronic. Customers with affected product were instructed to contact Medtronic Technical Services at 1-800-595-9709 to receive a return material authorization (RMA) and arrange for their return and no charge replacements. Customers were asked to complete and follow instructions on the attached Consignee Response Form. Sign and date the bottom of the form and then return the form to Medtronic at RS [email protected] or fax it to Medtronic Technical Services at 651-367-7075. Customers with questions were instructed to call 1-800-595-9709.
- Quantity in commerce
- 149 instruments
- Distribution
- Worldwide Distribution - US (nationwide) and Internationally to Canada.
- Product codes and lots
- Lot numbers 17030722, 17022421,17022020, 17020619, 17011918, 17012417, I 17010602, 16122724,16122239, 16122038, 16120563, and 16112362
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-3051-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-3052-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-3053-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-3054-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-3055-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-3056-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-3057-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-3058-2017 | Class II | Terminated | Voluntary: Firm initiated |