Z-3058-2017Class II

Medtronic ENT Prgm 9734636 Add Fusion Nav AxiEM, reusable, Rx. This kit includes 9733908 (which includes...

Medtronic Navigation, Inc / initiated 2017-06-29 / Terminated / root cause: design

Medtronic Navigation, Inc. began a recall of Medtronic ENT Prgm 9734636 Add Fusion Nav AxiEM, reusable, Rx. This kit includes 9733908 (which includes 9733452 containing 9733449, Straight Suction, 9733450... on . FDA classified it as Class II and gives the reason as "The instruments are not able to pass instrument verification which is performed prior to the use of each instrument". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-3058-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Navigation, Inc
Product
Medtronic ENT Prgm 9734636 Add Fusion Nav AxiEM, reusable, Rx. This kit includes 9733908 (which includes 9733452 containing 9733449, Straight Suction, 9733450, 70 deg Curved Suction, and 9733451, 90 deg Curved Suction). The firm name on the label is Medtronic Navigation, Inc., Louisville, CO. The suction devices are part of a family of reusable ENT instruments that are used with the StealthStationTM AxiEM surgical navigation system. The AxiEM system is an electromagnetic (EM) tracking solution for surgical tools. The system measures an induced voltage created by local magnetic fields from a set of coils in the transmit coil array (TCA). Each surgical tool has a set of receiving coils to sense the magnetic field strength. Using a mathematical algorithm, the position and orientation of the instrument can be computed and overlaid/registered to imagery on a computer monitor, resulting in the ability to track the instrument tip in relation to the patients anatomy. AxiEM ENT Suction devices are used as suction and pointing devices. The suction devices are intended to have an ENT Instrument Tracker mounted on it
Product code
PGW Ear, Nose, And Throat Stereotaxic Instrument
Reason for recall
The instruments are not able to pass instrument verification which is performed prior to the use of each instrument.
Root cause tag
design
FDA root cause
Device Design
Action
Medtronic sent an Urgent Medical Device Recall letter dated June 25, 2017, via overnight mail on June 29, 2017, to all affected customers. Letters to foreign customers are being handled by geography-specific Regulatory Specialists. Customers were instructed examine their inventory for the affected product and immediately quarantine them for return for Medtronic. Customers with affected product were instructed to contact Medtronic Technical Services at 1-800-595-9709 to receive a return material authorization (RMA) and arrange for their return and no charge replacements. Customers were asked to complete and follow instructions on the attached Consignee Response Form. Sign and date the bottom of the form and then return the form to Medtronic at RS [email protected] or fax it to Medtronic Technical Services at 651-367-7075. Customers with questions were instructed to call 1-800-595-9709.
Quantity in commerce
6 kits
Distribution
Worldwide Distribution - US (nationwide) and Internationally to Canada.
Product codes and lots
Each of the devices within the kit are labeled with information specific to each of the reusable devices within the kit and those devices are what goes into the customer inventory. The outer box where the lot number is printed is discarded by the customer on receipt and the separate devices within the box are what is stored on the shelf. Therefore, no kit lot number is being provided.To be provided

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-3058-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-3051-2017Class IITerminatedVoluntary: Firm initiated
Z-3052-2017Class IITerminatedVoluntary: Firm initiated
Z-3053-2017Class IITerminatedVoluntary: Firm initiated
Z-3054-2017Class IITerminatedVoluntary: Firm initiated
Z-3055-2017Class IITerminatedVoluntary: Firm initiated
Z-3056-2017Class IITerminatedVoluntary: Firm initiated
Z-3057-2017Class IITerminatedVoluntary: Firm initiated
Z-3058-2017Class IITerminatedVoluntary: Firm initiated