Z-0105-2018Class II

ROSA Brain 3.0 Intended for the spatial positioning and orientation of instrument holders or tool guides to...

Zimmer GmbH / initiated 2016-05-27 / Terminated / root cause: manufacturing process

Zimmer Biomet, Inc. began a recall of ROSA Brain 3.0 Intended for the spatial positioning and orientation of instrument holders or tool guides to be used by neurosurgeons to guide standard... on . FDA classified it as Class II and gives the reason as "Insufficient amount of washers in the Telescopic Arm, allowing some movement of the device". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0105-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
ROSA Brain 3.0 Intended for the spatial positioning and orientation of instrument holders or tool guides to be used by neurosurgeons to guide standard neurosurgical instruments. The device is indicated for any neurosurgical procedure in which the use of stereotactic surgery may be inappropriate.
Product code
HAW Neurological Stereotaxic Instrument
Reason for recall
Insufficient amount of washers in the Telescopic Arm, allowing some movement of the device.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Zimmer Biomet Field Service Technicians replaced Field Locking Systems on site.
Quantity in commerce
3 units
Distribution
Worldwide Distribution - US (DC) and Internationally to Australia and France.
Product codes and lots
Serial No. BR16004, BR15002, and BR15003

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0105-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0105-2018Class IITerminatedVoluntary: Firm initiated