Z-0115-2018Class II
ROSA Surgical Device 2.5.8
Zimmer GmbH / initiated 2016-04-08 / Terminated / root cause: software
Zimmer Biomet, Inc. began a recall of ROSA Surgical Device 2.5.8 on . FDA classified it as Class II and gives the reason as "Potential for software to change the final tool orientation for the command position without command". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0115-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- ROSA Surgical Device 2.5.8
- Product code
- HAW Neurological Stereotaxic Instrument
- Reason for recall
- Potential for software to change the final tool orientation for the command position without command.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Field Service Technicians visited consignee locations to perform system upgrades and have been completed.
- Quantity in commerce
- 64 units
- Distribution
- AL, AR, CA, CO, CT, FL, GA, KY, MA, MI, MO, NY, OH, PA, TX, and WA Canada, China, France, Germany, India, Israel, Italy, Russia, Saudi Arabia, Spain, and United Kingdom
- Product codes and lots
- Serial No. RO15050, RO15044, RO15048, RO15067, RO15045, RO15061, RO14041, RO14038, RO14039, RO14043, RO15051, RO15069, RO10011, RO14037, RO10014, RO15054, RO14035, RO15060, RO15063, RO14033, RO15059, RO15064, RO14031, RO14040, RO10009, RO13023, RO15053, RO15052, RO15058, RO13027, RO15046, and RO15062
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0115-2018 | Class II | Terminated | Voluntary: Firm initiated |