Z-0161-2018Class II

ROSA Surgical Device 2.5.8 It is intended to be used in the operating room for the spatial positioning and...

Zimmer GmbH / initiated 2014-06-04 / Terminated / root cause: software

Zimmer Biomet, Inc. began a recall of ROSA Surgical Device 2.5.8 It is intended to be used in the operating room for the spatial positioning and orientation of an instrument holder or tool guide... on . FDA classified it as Class II and gives the reason as "Software corrections reactivating the cooperative endoscopy mode". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0161-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
ROSA Surgical Device 2.5.8 It is intended to be used in the operating room for the spatial positioning and orientation of an instrument holder or tool guide. Guidance is based on a pre-operative plan developed with three-dimensional imaging software, and uses fiducial markers or optical registration. The system is intended for use by neurosurgeons to guide standard neurosurgical instruments. It is indicated for any neurosurgical condition in which the use of stereotactic surgery may be appropriate.
Product code
HAW Neurological Stereotaxic Instrument
Reason for recall
Software corrections reactivating the cooperative endoscopy mode.
Root cause tag
software
FDA root cause
Software Design Change
Action
The firm, Zimmer Biomet Inc., sent Field Service Technicians to consignee locations to perform system upgrade. All US units have been corrected. If you any question, please contact Post Market Surveillance and Regulatory Compliance Director at 574-372-4487 or email to: [email protected].
Quantity in commerce
29 devices(8 US and 21 OUS)
Distribution
Worldwide distribution: US (nationwide) in states of: AR, GA, MI, OH, and TX; and countries of:Canada, China, France, Germany, Israel, Italy, Russia, Saudi Arabia, and Spain.
Product codes and lots
Serial No. RO10011, RO14033, RO14031, RO14035, RO13027, RO13023, RO10014, RO10009, RO10010, RO14030, RO11016, RO11017, RO12022, RO14029, RO14032, RO08003, RO09004, RO10007, RO10013, RO12018, RO12019, RO13025, RO14034, RO11015, RO13024, RO09005, RO13026, RO12021, and RO12020.