Z-3041-2017Class II

HomeChoice Pro Automated PD system Model numbers: 5C8310, 5C8310R

Baxter Healthcare Corporation / initiated 2017-07-05 / Terminated / root cause: packaging

Baxter Healthcare Corporation began a recall of HomeChoice Pro Automated PD system Model numbers: 5C8310, 5C8310R Product Usage: The HomeChoice Automated Personal Cycler is intended for automatic control of... on . FDA classified it as Class II and gives the reason as "Baxter Healthcare Corporation has been made aware that users may not be following the instructions in the Operators Manual and incorrectly (using sharp objects) opening disposable set packaging while setting up their...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-3041-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Baxter Healthcare Corporation
Product
HomeChoice Pro Automated PD system Model numbers: 5C8310, 5C8310R Product Usage: The HomeChoice Automated Personal Cycler is intended for automatic control of dialysate solution exchanges in the treatment of pediatric and adult renal failure patients undergoing peritoneal dialysis.
Product code
FKX System, Peritoneal, Automatic Delivery
Reason for recall
Baxter Healthcare Corporation has been made aware that users may not be following the instructions in the Operators Manual and incorrectly (using sharp objects) opening disposable set packaging while setting up their Peritoneal Dialysis (PD) therapy, damaging the cassettes for the HomeChoice or HomeChoice PRO cyclers. If damaged cassettes are used, the cyclers may not consistently detect very small holes/cuts in the sheeting of the cassette in the patient valve region and the cycler may deliver air into the patient.
Root cause tag
packaging
FDA root cause
Use error
Action
The firm notified their consignees by letter on 07/05/2017 of the potential for damaging the disposable set if tools are used to open the packaging and directing them to only open the packaging by hand by letter.
Quantity in commerce
23732 units
Distribution
Worldwide Distribution
Product codes and lots
all codes

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-3041-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-3040-2017Class IITerminatedVoluntary: Firm initiated
Z-3041-2017Class IITerminatedVoluntary: Firm initiated