510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0291-2018Class II

Orthofix Power Drill Torque Limiter, Sterile, Rx Only, Single Use; Sold individually and in the Galaxy UNYCO...

Orthofix SRL / initiated 2017-07-05 / Terminated

Orthofix Srl began a recall of Orthofix Power Drill Torque Limiter, Sterile, Rx Only, Single Use; Sold individually and in the Galaxy UNYCO Diaphyseal Tibia Sterile Kit, Galaxy UNYCO Mini... on . FDA classified it as Class II and gives the reason as "The power drill torque limiter may break during use in surgery application". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0291-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Orthofix SRL
Product
Orthofix Power Drill Torque Limiter, Sterile, Rx Only, Single Use; Sold individually and in the Galaxy UNYCO Diaphyseal Tibia Sterile Kit, Galaxy UNYCO Mini Kit Instruments Sterile, Galaxy UNYCO Diaphyseal Femur Box, Galaxy UNYCO Knee Bridging Box, Galaxy UNYCO Ankle Bridging Box - Unilateral Frame, Galaxy UNYCO Ankle Bridging Box - Delta Frame, and Galaxy UNYCO Diaphyseal Tibia Box. Intended to be used for temporary bone stabilization in trauma and orthopedic procedures of the lower limb prior to definitive treatment.
Product code
KTT Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component
Reason for recall
The power drill torque limiter may break during use in surgery application.
FDA root cause
Process change control
Action
Consignees were sent an Urgent Medical Device Recall Notification, dated June 28, 2017, informing them of the device correction. The letter instructed distributors to immediately identify affected product in warehouses, quarantine affected product, and communicate the amount on hand to Orthofix SRL for replacement. Any customers are to be forwarded the recall notice provided to the distributor. Distributors are asked to complete the Reply Form with the appropriate information and to return it to Orthofix by fax at +39-045-6719380 or by e-mail to [email protected]. Hospitals were instructed to identify affected product on-site, quarantine affected units, and communicate the quantity of affected product on-hand to their local distributor or to Orthofix. Affected product will be replaced by Orthofix.
Quantity in commerce
416 units
Distribution
Worldwide Distribution -- US, including the states of CA, FL, & PA; and, the countries of Australia, Germany, Finland, France, Italy, Lithuania, Netherlands, Singapore, South Africa, Sweden, Switzerland, & United Kingdom.
Product codes and lots
Power Drill Torque Limiter: Product Code 99-93568; Lot (Exp Date) V1388296 (12/28/2019), V1388432 (12/29/2019), V1389857 (01/13/2020), V1400707 (05/18/2020), V1400940 (05/23/2020), V1410027 (09/01/2020); UDI: 18054242512318.   Galaxy UNYCO Diaphyseal Tibia Sterile Kit: Product Code 99-93506; Lot (Exp Date) V1389818 (01/12/2020), V1389819 (01/12/2020), V1389886 (02/13/2020), V1399881 (05/10/2020), V1400800 (05/22/2020), V1402023 (06/03/2020), V1405602 (07/17/2020), V1406422 (07/26/2020), V1410082 (09/02/2020), R1412578 (09/28/2020), V1414451 (10/19/2020), V1421785 (02/01/2021), V1425406 (03/13/2021), B1012462 (06/19/2021), B1012776 (06/26/2021), B1019431 (07/05/2021), B1031608 (08/05/2021), B1052165 (10/23/2021), B1049195 (10/25/2021), B1080028 (02/16/2022), B1099157 (04/08/2022), B1107271 (04/09/2022); UDI: 18054242512325.   Galaxy UNYCO Mini Kit Instruments Sterile: Product Code 99-93509; Lot (Exp Date) V1408711 (08/01/2020), V1410026 (09/01/2020), V1422854 (02/14/2021), V1427350 (04/09/2021), V1429217 (04/26/2021), B1083360 (02/20/2022); UDI: 18054242513629.   Galaxy UNYCO Diaphyseal Femur Box: Product Code 99-93790; Lot (Exp Date) B1019281 (04/09/2021); UDI: 1805424515159.   Galaxy UNYCO Knee Bridging Box: Product Code 99-93791; Lot (Exp Date) B1019282 (04/09/2021); UDI: 18054242515166.  Galaxy UNYCO Ankle Bridging Box - Unilateral Frame; Product Code 99-93792; Lot (Exp Date) V140597 (03/21/2020); UDI: 18054242513643.   Galaxy UNYCO Ankle Bridging Box - Delta Frame: Product Code 99-93793; Lot (Exp Date) V1409944 (03/09/2020); UDI: 18054242513650.   Galaxy UNYCO Diaphyseal Tibia Box: Product Code 99-93794; Lot (Exp Date) V1410194 (04/18/2020). UDI: 18054242513667.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0291-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0291-2018Class IITerminatedVoluntary: Firm initiated