510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-3042-2017Class II

Arctic Sun 5000 Temperature Management System, 100-120 V North America and 100-120 V US

C.R. Bard Inc / initiated 2017-03-31 / Terminated

C.R. Bard, Inc. began a recall of Arctic Sun 5000 Temperature Management System, 100-120 V North America and 100-120 V US on . FDA classified it as Class II and gives the reason as "Potential failure of the Arctic Sun 5000 Temperature Management System may occur when the continued use is in excess of 2000 system hours without preventive maintenance". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-3042-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
C.R. Bard Inc
Product
Arctic Sun 5000 Temperature Management System, 100-120 V North America and 100-120 V US
Product code
DWJ System, Thermal Regulating
Reason for recall
Potential failure of the Arctic Sun 5000 Temperature Management System may occur when the continued use is in excess of 2000 system hours without preventive maintenance.
FDA root cause
Unknown/Undetermined by firm
Action
Bard sent an Urgent Customer Notification letter dated March 31, 2017, to each of the U.S. consignees. The letter contained recommended actions and included an Acknowledgement Form to be returned. For further questions, please call (770) 784-6119.
Quantity in commerce
3,450 units
Distribution
Worldwide Distribution - USA (nationwide) to the states of :AL, AK, AZ, AR, CA, CO, CT, DE, DC, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WY., and to the countries of : Australia, Korea, Japan, Canada, Hong Kong, China, Singapore, & EMEA
Product codes and lots
Product Codes 5000-00-00 & 5000-00-00E - All serial numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-3042-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-3042-2017Class IIOngoingVoluntary: Firm initiated