VITEK 2 Compact 15, REF numbers: 27415 and 27415R software and reagent cards designed for the identification...
Biomerieux Inc began a recall of VITEK 2 Compact 15, REF numbers: 27415 and 27415R software and reagent cards designed for the identification (ID) and antimicrobial susceptibility testing... on . FDA classified it as Class II and gives the reason as "Customers have reported that some VITEK 2 cards are staying in preliminary status, not finalizing after ejection from the instrument, and not allowing cards in subsequent carousel slots to be processed. The issue was...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-3075-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- bioMerieux, Inc.
- Product
- VITEK 2 Compact 15, REF numbers: 27415 and 27415R software and reagent cards designed for the identification (ID) and antimicrobial susceptibility testing (AST) of bacteria and yeast.
- Product code
- LON System, Test, Automated, Antimicrobial Susceptibility, Short Incubation
- Reason for recall
- Customers have reported that some VITEK 2 cards are staying in preliminary status, not finalizing after ejection from the instrument, and not allowing cards in subsequent carousel slots to be processed. The issue was reported to occur on VITEK 2 Compact 15 and Compact 30 systems following a system software update to version 8.01.
- Root cause tag
- software
- FDA root cause
- Other
- Action
- Consignees (Subsidiaries) of bioMrieux, impacted by the identified issue, were notified of the FSCA and corrective action (FSCA notification) electronically on 07/12/2017. The direct consignees (subsidiaries and distributors) have until 08/12/2017 to notify their respective impacted users. The directions to the user are as follows: Please implement the following actions at this time: " Confirm this letter has been distributed to, and reviewed by, all appropriate personnel within your organization. " Refrain from performing the VITEK 2 Systems Software 8.01 update. o A solution has been identified, and your local bioMrieux representative will contact you with further instruction in the coming weeks. " If you have updated VITEK 2 Systems Software to version 8.01 within the last seven (7) days, please contact your local bioMrieux representative. " Please store this letter with your bioMrieux instrument documentation. " Complete the Acknowledgement Form and return it to your local bioMrieux representative."
- Quantity in commerce
- 1866 units
- Distribution
- Worldwide
- Product codes and lots
- VITEK¿ 2 Compact 15 and Compact 30 instruments previously manufactured/installed with VITEK¿ 2 System software versions prior to 5.01, and currently being upgraded to version 8.01.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-3075-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-3076-2017 | Class II | Terminated | Voluntary: Firm initiated |