510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0064-2018Class II

Cobalt HV Bone Cement 40GM, REF 402282, QTY 1, STERILE, Rx Only, Mfgr: BIOMET ORTHOPEDICS Cobalt MV and HV...

Encore Medical, LP / initiated 2017-06-26 / Open, Classified / root cause: supplier

Encore Medical, Lp began a recall of Cobalt HV Bone Cement 40GM, REF 402282, QTY 1, STERILE, Rx Only, Mfgr: BIOMET ORTHOPEDICS Cobalt MV and HV Bone Cement is indicated for use as bone cement in... on . FDA classified it as Class II and gives the reason as "Loss of the seal on the sterile Tyvek packaging used with this Cobalt Bone Cement". As of , the recall status on record is Open, Classified, and MDM tags the root cause as supplier.

Recall record

Recall number
Z-0064-2018
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Encore Medical, LP
Product
Cobalt HV Bone Cement 40GM, REF 402282, QTY 1, STERILE, Rx Only, Mfgr: BIOMET ORTHOPEDICS Cobalt MV and HV Bone Cement is indicated for use as bone cement in arthroplastic procedures of the hip, knee and other joints to fix plastic and metal prosthetic parts to living bone when reconstructions is necessary because of osteoarthritis, rheumatoid arthritis, traumatic arthritis, avascular necrosis, nonunion of fractures of the neck of the femur, sickle cell anemia osteoporosis, secondary severe joint destruction following trauma or other conditions (also far fixation of unstable fractures in metastatic malignancies), and revision of previous arthroplasty procedures
Product code
LOD Bone Cement
Reason for recall
Loss of the seal on the sterile Tyvek packaging used with this Cobalt Bone Cement.
Root cause tag
supplier
FDA root cause
Vendor change control
Action
Zimmer Biomet sent an Urgent Medical Device Recall Notice dated April 2017 to all affected customers. The firm initiated their recall to their distributors on 06/26/2017 requesting that they destroy any product on hand. The firm then expanded their recall on 09/14/2017, beginning their notification by email and following with letters to hospitals delivered on 09/21 and 09/22/2017. The firm requested the following actions: "1. Complete the Acknowledgement/Questionnaire provided with this lot specific field safety notice and return to Stericycle by email to [email protected] or by fax to 866-608-3939. If you have any questions or issues related to sending this information, please call 877-551-7153. 2. Quarantine this affected product. Arrangements will be made to finalize the handling of the product as well as coordinate the replacement product where applicable. 3. If packaging issues are found, please contact DJO Surgical Customer Service to initiate a product complaint." Customers with questions were instructed to call 512-834-6255.
Quantity in commerce
7143 units
Distribution
Nationwide Distribution
Product codes and lots
Lot Numbers: 508220, 189680, 199200, 310390, 424730, 668130, 668150, 871240, 959720

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0064-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0064-2018Class IIOngoingVoluntary: Firm initiated
Z-0065-2018Class IIOngoingVoluntary: Firm initiated
Z-0066-2018Class IIOngoingVoluntary: Firm initiated
Z-0067-2018Class IIOngoingVoluntary: Firm initiated