Z-3077-2017Class II
Fogarty Adherent Clot Catheter Model Numbers 1408010 and 140806
Edwards Lifesciences LLC / initiated 2017-07-06 / Terminated / root cause: packaging
Edwards Lifesciences, LLC began a recall of Fogarty Adherent Clot Catheter Model Numbers 1408010 and 140806 Product Usage: The Fogarty adherent clot catheters are designed to effectively remove clot... on . FDA classified it as Class II and gives the reason as "The diameter of the Fogarty Catheter balloon, measures 6mm instead of 10mm". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-3077-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Edwards Lifesciences LLC
- Product
- Fogarty Adherent Clot Catheter Model Numbers 1408010 and 140806 Product Usage: The Fogarty adherent clot catheters are designed to effectively remove clot material in the peripheral vasculature. The catheters are indicated for the removal of emboli and thrombi from native arteries or synthetic grafts. The device are designed to facilitate clots too resistant to be removed by an elastomeric balloon. The device features a spiral-shaped, latex-covered stainless steel cable that assumes a corkscrew shape when retracted, greatly expanding the surface areas to entrap fibrous material. The device is 80 cm in length and is available in 4F-6F diameters.
- Product code
- DXE Catheter, Embolectomy
- Reason for recall
- The diameter of the Fogarty Catheter balloon, measures 6mm instead of 10mm.
- Root cause tag
- packaging
- FDA root cause
- Packaging
- Action
- Recall notification letters were sent to affected customers on 7/6/2017. The recall notice included a response form and instructed consignee to return any unused devices..
- Quantity in commerce
- 200 units
- Distribution
- Worldwide Distribution - US Nationwide GB, NL, DE, IT, FR, BE, ES, NO, CZ, BG, SE, AT, CH, PL, RU, FI
- Product codes and lots
- Lots 60439817 and 60439818
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-3077-2017 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY