SORIN GROUP, SMART PERFUSION PACK, SMA PRIMO2X OPTI PACK, REF 088512900, STERILE EO, Rx Only, 1 EA
Sorin Group USA, Inc. began a recall of SORIN GROUP, SMART PERFUSION PACK, SMA PRIMO2X OPTI PACK, REF 088512900, STERILE EO, Rx Only, 1 EA on . FDA classified it as Class II and gives the reason as "Medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter). The result is a potential sterile...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-2785-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- LivaNova USA, Inc.
- Product
- SORIN GROUP, SMART PERFUSION PACK, SMA PRIMO2X OPTI PACK, REF 088512900, STERILE EO, Rx Only, 1 EA
- Product code
- DWE Tubing, Pump, Cardiopulmonary Bypass
- Reason for recall
- Medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter). The result is a potential sterile barrier breach in the display sets. These display sets are included with LivaNova Perfusion Tube Sets (PTS) which were then distributed to customers.
- Root cause tag
- packaging
- FDA root cause
- Other
- Action
- The firm initiated their recall verbally on 06/29/2017 and following with letters mailed on 07/13/2017. The actions to be taken by the customer: 1. Notify staff to not use the Medtronic labeled device contained within the LivaNova Perfusion Tube Set packs. 2. If the set is a standalone component within the pack, remove the device upon opening the pack at the time of the procedure. 3. If the device is connected to the circuit, remove the device and attach a new unit upon opening the pack. 4. Identify the Medtronic device with one of the barcode pack labels attached to the lidstock of the LivaNova perfusion pack. 5. Collect the devices for return to LivaNova until all affected perfusion sets have been consumed. In order to assist in controlling inventory, the LivaNova sales representative will have labels available to identify packs containing the affected devices. 6. Contact LivaNova Customer Support at (800) 650-2623 (M-F 8AM to 5PM MT) to arrange for the return of the affected product, and to order equivalent replacement product.
- Quantity in commerce
- 524 units
- Distribution
- Nationwide
- Product codes and lots
- Lot Numbers: 1620700048, 1621600063, 1623700039, 1624500038, 1625900047, 1626500084, 1627800066, 1628500062, 1630000040, 1630500055, 1631400035, 1633300052, 1635400049, 1700600038, 1702500069, 1704500039, 1706600107, 1707400082, 1708600072, 1710800102, 1711700034, 1713600045, 1715000038, 1717000046
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2783-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2784-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2785-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2786-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2787-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2788-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2789-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2790-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2791-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2792-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2793-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2794-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2795-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2796-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2797-2017 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2798-2017 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.