Z-3043-2017Class II

Encore(TM) 26 Inflation Device, REF Catalog Nos.: (a) H74904526011, (b) M001151050, (c) M00566670, (d)...

Boston Scientific Corporation / initiated 2017-06-15 / Terminated

Boston Scientific Corporation began a recall of Encore(TM) 26 Inflation Device, REF Catalog Nos.: (a) H74904526011, (b) M001151050, (c) M00566670, (d) M0067101140 on . FDA classified it as Class II and gives the reason as "Boston Scientific (BSC) has received complaints for cracked trays on the Encore device". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-3043-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
Encore(TM) 26 Inflation Device, REF Catalog Nos.: (a) H74904526011, (b) M001151050, (c) M00566670, (d) M0067101140
Product code
MAV Syringe, Balloon Inflation
Reason for recall
Boston Scientific (BSC) has received complaints for cracked trays on the Encore device.
FDA root cause
Under Investigation by firm
Action
The firm initiated their recall on 06/15/2017 by letter delivered by Fed Ex in the US, on 06/15-29/2017 in the Americas delivered by Courier, email and hand delivery, on 06/16/2017 in Europe by DHL, on 06/16-19/2017 in the Arab Emirates and East Asia by email and hand delivery and on 06/136/2017 in Japan by hand delivery. In the notice, the firm described the problem and requested the following action: 1. Immediately check the inventory and identify whether there are products with UPNs and lot numbers affected by the recall. 2. Inspect the device tray in the area highlighted above. If any units with a cracked tray are identified in the supply, immediately discontinue use and segregate affected units in a secure location. 3. Arrange for the damaged units to be returned to Boston Scientific by completing the attached Reply Verification Tracking Form. 4. If no damage is identified, the device may be used as normal. Please complete and return the attached Reply Verification Tracking Form
Quantity in commerce
305,779 units
Distribution
Worldwide
Product codes and lots
Serial Numbers: (a) 19563006, 19587644, 19623142, 19650458, 19704550, 19725489, 19769215, 20054530, 20077271, 20083051, 20127928, 20159518, 20202728, 20206930, 20242335, 20285571, 20309573, 20309574, 20336380, 20394218, 20423644, 20423645, 20456071, 20481170, 20534557  (b) 19640750, 19667072, 19704551, 19755302, 19755307, 19785021, 19808183, 20030825, 20060033, 20087896, 20117909, 20159601, 20171449, 20206929, 20233700, 20256035, 20279769, 20303751, 20336385, 20336386, 20364928, 20364929, 20364930, 20423650, 20472532, 20493563, 20507373, 20534581  (c) 20157078, 19623019, 19755304, 20056496, 20146417, 20264032, 20456079  (d) 19594194, 19717352, 19808180, 20121252, 20223176, 20336387, 20394350, 20534582

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-3043-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-3043-2017Class IITerminatedVoluntary: Firm initiated