ORA System with VerifEye, Catalog Number 8065998300 For use during intraocular lens surgery

Alcon Laboratories / initiated 2017-06-30 / Terminated / root cause: software

Alcon Research, Ltd. began a recall of ORA System with VerifEye, Catalog Number 8065998300 For use during intraocular lens surgery on . FDA classified it as Class II and gives the reason as "Some ORA Carts have the potential to return an incorrect IOL power measurement during cataract surgery. This issue appears to have been caused by a software coding error that results in the lens coefficients for an IOL...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-3050-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Alcon Laboratories
Product
ORA System with VerifEye, Catalog Number 8065998300 For use during intraocular lens surgery
Product code
NCF Aberrometer, Ophthalmic
Reason for recall
Some ORA Carts have the potential to return an incorrect IOL power measurement during cataract surgery. This issue appears to have been caused by a software coding error that results in the lens coefficients for an IOL model being downloaded from the Alcon server in an incorrect order.
Root cause tag
software
FDA root cause
Software design
Action
The Market Action was initiated with Initial telephone contact to all affected customers starting June 30, 2017. These customers were informed of the issue, and advised not to use their ORA Carts for calculations with the lens model affected on their specific system. A confirmatory letter (Attachment 4) was also sent via overnight mail to these customers July 12, 2017. As of July 12, 2017, Alcon has identified 8 ORA Carts that are affected by the software coding error. Alcon has reset the IOL databases on all 8 of the identified affected ORA Carts.
Quantity in commerce
367 units
Distribution
Worldwide distribution. US Nationwide and Argentina, Australia, Belgium, Brazil, Canada, China, Colombia, France, India, Japan, Mexico, Netherlands, Panama, Portugal, Romania, Spain, Thailand, UAE, and United Kingdom
Product codes and lots
Serial Numbers: C0002,C2006,C2020,C2026,C2054,C3009R,C3010R,C3024,C3036,C3066,C3067,C3085R,C3090,C3093,C3108,C3136,C3153,C4001,C4003,C4005,C4006,C4007,C4009,C4010,C4011,C4012,C4013,C4014,C4015,C4016,C4017,C4019,C4020,C4021,C4022,C4023,C4024,C4025,C4026,C4027,C4028,C4029,C4030,C4031,C4035,C4036,C4037,C4038,C4039,C4040,C4043,C4045,C4046R,C4047R,C4049R,C4050R,C4052R,C4053R,C4054R,C4055R,C4056R,C4057R,C4058R,C4059R,C4060R,C4061R,C4062R,C4065R,C4066R,C4067R,C4068R,C4069R,C4070R,C4071R,C4072R,C4073R,C4074R,C4075R,C4076R,C4077R,C4078R,C4079R,C4080R,C4083R,C4085R,C4086R,C4087RC4088R,C4089R,C4090R,C4091R,C4092R,C4093R,C4094R,C4095R,C4096R,C4097R,C4098R,C4099R,C4100R,C4101R,C4102R,C4103R,C4104R,C4105R,C4106R,C4107R,C4108R,C4109R,C4110R,C4111R,C4112R,C4113R,C4115R,C4117R,C4118R,C4119R,C4120R,C4121R,C4122R,C4123R,C4124R,C4126R,C4127R,C4128R,C4129R,C4130R,C4132R,C4133R,C4134R,C4135R,C4136R,C4138R,C4140,C4142,C4143,C4144,C4145,C4146,C4147,C4148,C4149,C4150,C4151,C4152R,C4153,C4154,C4155,C4156,C4157R,C4158,C4159,C4160R,C4161R,C4162R,C4163R,C4164R,C4165R,C4166R,C4167R,C4168R,C4169R,C4170R,C4173R,C4174R,C4175,C4177R,C4178R,C4179R,C4180R,C4181R,C4182R,C4183R,C4184R,C4185R,C4186R,C4187R,C4188R,C4189R,C4190R,C4191R,C4192R,C4193R,C4195R,C4196R,C4197R,C4198,C4199R,C4200R,C4201,C4202R,C4203R,C4204R,C4205R,C4206R,C4207R,C4208R,C4209R,C4210R,C4211R,C4212R,C4213R,C4214R,C4215R,C4216R,C4217R,C4218R,C4219R,C4220R,C4221R,C4222R,C4223R,C4224R,C4225R,C4226R,C4227R,C4228R,C4229R,C4230R,C4231R,C4232R,C4233R,C4234R,C4235R,C4238R,C4239R,C4240R,C4241R,C4242R,C4243R,C4244R,C4245R,C4246R,C4247R,C4248R,C4249R,C4251R,C4252R,C4253R,C4254R,C4255R,C4256R,C4257R,C4258R,C4259R,C4260R,C4261R,C4262R,C4264R,C4266R,C4267R,C4268R,C4269R,C4271R,C4272R,C4273R,C4274R,C4275R,C4276R,C4278R,C4279R,C4280R,C4283R,C4284R,C4285R,C4286R,C4287R,C4288R,C4289R,C4290R,C4291R,C4292R,C4293R,C4294R,C4295R,C4296R,C4297R,C4298R,C4299R,C4300R,C4301R,C4302R,C4303R,C4304R,C4305R,C4306R,C4307R,C4308R,C4309R,C4310R,C4311R,C4312R,C4315R,C4316R,C4317R,C4318R,C4319R,C4320R,C4321R,C4323R,C4324R,C4326R,C4327R,C4328R,C4331R,C4332R,C4334R,C4337R,C4339R,C4340R,C4341R,C4342R,C4343R,C4344R,C4345R,C4346R,C4347R,C4348R,C4349R,C4352R,C4354R,C4355R,C4356R,C4357R,C4358R,C4359R,C4360R,C4361R,C4363R,C4364R,C4365R,C4366R,C4367R,C4368R,C4371R,C4372R,C4373R,C4374R,C4376R,C4377R,C4378R,C4379R,C4380R,C4381R,C4382R,C4383R,C4384R,C4385R,C4387R,C4388R,C4389R,C4394,C4396,C4397,C4399,C4406R,C4407R,C4408R,C4415,C4418,C4420,C4421,C4422,C4423,C4424,C4425,C4426

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-3050-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-3049-2017Class IITerminatedVoluntary: Firm initiated
Z-3050-2017Class IITerminatedVoluntary: Firm initiated