Z-3100-2017Class II

smith&nephew POLARSTEM(TM) Femoral Stems with Ti/HA (Standard, Lateral, Valgus and Collar), REF 76102260, NON...

Smith & Nephew, Inc. / initiated 2017-07-24 / Terminated / root cause: design

Smith & Nephew, Inc. began a recall of smith&nephew POLARSTEM(TM) Femoral Stems with Ti/HA (Standard, Lateral, Valgus and Collar), REF 76102260, NON STERILE, QTY: (1), Rx only The subject cardan is... on . FDA classified it as Class II and gives the reason as "The Cardan joint of the Polarstem Cardan could potentially fracture or functionally fail after multiple procedures". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-3100-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Smith & Nephew, Inc.
Product
smith&nephew POLARSTEM(TM) Femoral Stems with Ti/HA (Standard, Lateral, Valgus and Collar), REF 76102260, NON STERILE, QTY: (1), Rx only The subject cardan is used in conjunction with a curved stem impactor used to implant the POLARSTEM hip stem. The hip stem is mounted on the screw at the tip of the instrument, and the stem is then impacted into the femur. After impaction, the surgeon uses a knob at the end of the cardan to unscrew the implanted stem from the instrument
Product code
LZO Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented
Reason for recall
The Cardan joint of the Polarstem Cardan could potentially fracture or functionally fail after multiple procedures.
Root cause tag
design
FDA root cause
Device Design
Action
Smith & Nephew sent an Urgent Medical Device Recall Notice dated July 24, 2017, to all affected consignees. The firm notified their consignees by email and Fed Ex on 07/24/2017. Consignees have been asked to inspect inventory to locate and quarantine affected product. All consignees were asked to complete a Response Form provided with the Field Safety Notice, regardless whether they have affected product. Affected inventory is then to be returned back to Smith & Nephew. Customers with questions were instructed to contact [email protected]. For questions regarding this recall call 978-749-1440.
Quantity in commerce
402 units
Distribution
Worldwide Distribution - US (Nationwide) and Internationally to Australia, Belgium, Norway, and Switzerland.
Product codes and lots
Lot Numbers: A57835, A57837, A58509, A59084, A59446, A59547, A60948

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-3100-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-3100-2017Class IITerminatedVoluntary: Firm initiated