Z-0087-2018Class II
RX-2 COUDE Epidural Needle, 16g x 3.5in (8.9cm)TW, 10 units, Calibrated, Plastic Hub, Detachable Wing, Blunt...
Epimed International, Inc. / initiated 2017-07-24 / Terminated / root cause: manufacturing process
Epimed International began a recall of RX-2 COUDE Epidural Needle, 16g x 3.5in (8.9cm)TW, 10 units, Calibrated, Plastic Hub, Detachable Wing, Blunt Stylet, Sterile, Rx only, For Use with 19g or... on . FDA classified it as Class II and gives the reason as "Potential for partially incomplete seal of product, impacting sterility". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0087-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Epimed International, Inc.
- Product
- RX-2 COUDE Epidural Needle, 16g x 3.5in (8.9cm)TW, 10 units, Calibrated, Plastic Hub, Detachable Wing, Blunt Stylet, Sterile, Rx only, For Use with 19g or Smaller Catheter
- Product code
- OGE Epidural Anesthesia Kit
- Reason for recall
- Potential for partially incomplete seal of product, impacting sterility.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- Consignees were contacted 07/24/2017 via email and phone, notifying them of the recall. Consignees were asked to contact the recalling firm to report the number of affected product on hand. A second, updated recall notification was sent on 08/04/2017. Instructions included to examine inventory and quarantine any affected product, contact Epimed to arrange return/credit of the affected product, and complete and return the response form.
- Quantity in commerce
- 500 units
- Distribution
- Domestic: AR, CA, IA, IN, KY, LA, MI, NY, OH, OR, TN, TX, & WA. Foreign: Canada, Czech Republic, Germany ,& Turkey.
- Product codes and lots
- Lot No. 12168200; Ref No. 107-1416.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0087-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0088-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0089-2018 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.