Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0088-2018Class II

COUDE Blunt Nerve Block Needle, 10 units, 0.71mm (22 Ga) x 152mm, Each Unit Contains One Blunt Nerve Block...

Epimed International, Inc. / initiated 2017-07-24 / Terminated / root cause: manufacturing process

Epimed International began a recall of COUDE Blunt Nerve Block Needle, 10 units, 0.71mm (22 Ga) x 152mm, Each Unit Contains One Blunt Nerve Block Needle and One Wing, Sterile on . FDA classified it as Class II and gives the reason as "Potential for partially incomplete seal of product, impacting sterility". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0088-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Epimed International, Inc.
Product
COUDE Blunt Nerve Block Needle, 10 units, 0.71mm (22 Ga) x 152mm, Each Unit Contains One Blunt Nerve Block Needle and One Wing, Sterile
Product code
OGE Epidural Anesthesia Kit
Reason for recall
Potential for partially incomplete seal of product, impacting sterility.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
Consignees were contacted 07/24/2017 via email and phone, notifying them of the recall. Consignees were asked to contact the recalling firm to report the number of affected product on hand. A second, updated recall notification was sent on 08/04/2017. Instructions included to examine inventory and quarantine any affected product, contact Epimed to arrange return/credit of the affected product, and complete and return the response form.
Quantity in commerce
300 units
Distribution
Domestic: AR, CA, IA, IN, KY, LA, MI, NY, OH, OR, TN, TX, & WA. Foreign: Canada, Czech Republic, Germany ,& Turkey.
Product codes and lots
Lot No. 12608201; Ref No. 117-2260.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0088-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0087-2018Class IITerminatedVoluntary: Firm initiated
Z-0088-2018Class IITerminatedVoluntary: Firm initiated
Z-0089-2018Class IITerminatedVoluntary: Firm initiated