510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-3127-2017Class II

Firebird NXG Spinal Fixation System Decorticating Planer, Catalog #36-1334, non-sterile, Rx. .A Decorticating...

Orthofix SRL / initiated 2017-08-02 / Terminated / root cause: manufacturing process

Orthofix, Inc began a recall of Firebird NXG Spinal Fixation System Decorticating Planer, Catalog #36-1334, non-sterile, Rx. .A Decorticating Planer is used to decorticate bone surrounding... on . FDA classified it as Class II and gives the reason as "The plastic core of the Decorticating Planer was found to be missing, which could result in scoring of the bone screw neck if used at or beyond a 23 degree angle". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-3127-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Orthofix SRL
Product
Firebird NXG Spinal Fixation System Decorticating Planer, Catalog #36-1334, non-sterile, Rx. .A Decorticating Planer is used to decorticate bone surrounding the spherical head of an implanted bone screw in order to provide clearance to allow for proper seating of the modular body, as well as to provide full range of modular body angulation. It is placed axially over the spherical head of the bone screw and rotated clockwise and counterclockwise to achieve the needed decortication
Product code
NKB Thoracolumbosacral Pedicle Screw System
Reason for recall
The plastic core of the Decorticating Planer was found to be missing, which could result in scoring of the bone screw neck if used at or beyond a 23 degree angle.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Orthofix sent an Urgent Medical Device Product Recall letter dated August 2, 2017, to all affected customers. The letter explained the issue and requested return of product in their possession. Customers with questions were instructed to contact their local Orthofix representative. For questions regarding this recall call 214-937-2100.
Quantity in commerce
6 planers
Distribution
Distribution was made to distributors located in CA, FL, and RI. There was no foreign/military/government distribution.
Product codes and lots
Lot #203013-TD26

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-3127-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-3127-2017Class IITerminatedVoluntary: Firm initiated