Firebird NXG Spinal Fixation System Decorticating Planer, Catalog #36-1334, non-sterile, Rx. .A Decorticating...
Orthofix, Inc began a recall of Firebird NXG Spinal Fixation System Decorticating Planer, Catalog #36-1334, non-sterile, Rx. .A Decorticating Planer is used to decorticate bone surrounding... on . FDA classified it as Class II and gives the reason as "The plastic core of the Decorticating Planer was found to be missing, which could result in scoring of the bone screw neck if used at or beyond a 23 degree angle". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-3127-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Orthofix SRL
- Product
- Firebird NXG Spinal Fixation System Decorticating Planer, Catalog #36-1334, non-sterile, Rx. .A Decorticating Planer is used to decorticate bone surrounding the spherical head of an implanted bone screw in order to provide clearance to allow for proper seating of the modular body, as well as to provide full range of modular body angulation. It is placed axially over the spherical head of the bone screw and rotated clockwise and counterclockwise to achieve the needed decortication
- Product code
- NKB Thoracolumbosacral Pedicle Screw System
- Reason for recall
- The plastic core of the Decorticating Planer was found to be missing, which could result in scoring of the bone screw neck if used at or beyond a 23 degree angle.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Orthofix sent an Urgent Medical Device Product Recall letter dated August 2, 2017, to all affected customers. The letter explained the issue and requested return of product in their possession. Customers with questions were instructed to contact their local Orthofix representative. For questions regarding this recall call 214-937-2100.
- Quantity in commerce
- 6 planers
- Distribution
- Distribution was made to distributors located in CA, FL, and RI. There was no foreign/military/government distribution.
- Product codes and lots
- Lot #203013-TD26
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-3127-2017 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY