Z-3142-2017Class II

Spine Referencing Instrumentation, Double Spinous Process Clamp, Short When used with a Medtronic...

Medtronic Navigation, Inc / initiated 2017-06-29 / Terminated / root cause: design

Medtronic Navigation, Inc. began a recall of Spine Referencing Instrumentation, Double Spinous Process Clamp, Short When used with a Medtronic StealthStation Navigation System, the Spine Referencing... on . FDA classified it as Class II and gives the reason as "Potential for Spinal Clamps to be damaged when opened beyond intended limits, either no opening once attached to a patient's spinous process or leaving a component behind in the patient's body following breakage". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-3142-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Navigation, Inc
Product
Spine Referencing Instrumentation, Double Spinous Process Clamp, Short When used with a Medtronic StealthStation Navigation System, the Spine Referencing fixation devices are intended to provide rigid attachment between patient and patient reference frame for the duration of the surgery. The devices are intended to be reusable. The spine referencing clamps provide bony fixation through two styles of clamps; a single spinous process clamp and a double spinous process clamp.
Product code
OLO Orthopedic Stereotaxic Instrument
Reason for recall
Potential for Spinal Clamps to be damaged when opened beyond intended limits, either no opening once attached to a patient's spinous process or leaving a component behind in the patient's body following breakage.
Root cause tag
design
FDA root cause
Device Design
Action
Medtronic sent an Urgent Medical Device Correction letter dated June 27, 2017, to all affected consignees. The letter notifies consignees of the product correction and requests that practitioners follow the attached use instructions for devices. Practitioners are to identify product damage before each procedure by opening and closing the device jaw with the T-handle. It is instructed that practitioners not continue turning the screw if resistance is met, as this may break the washer. Clamps are not to be used if the jaws remain closed when attempting to open a clamp or if the clamp screw back is out of the clamp. It is asked that this check is done to all inventory and any damaged product is to be reported to Medtronic at 1-800-595-9709 and quarantined. Revised product is available. Consignees are asked to complete and return the provided Consignee Response Form by email to the firm at [email protected] or by fax to 651-367-7075. Any questions can be directed to 1-800-595-9709.
Quantity in commerce
2,051 units total
Distribution
Worldwide Distribution - US including AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, AD, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, WV, & WY. Foreign: Albania, Australia, Bangladesh, Belgium, Bulgaria, Canada, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Indonesia, Israel, Italy, Japan, Kenya, Latvia, Luxembourg, Malaysia, Martinique, Netherlands, New Zealand, Philippines, Poland, Portugal, Romania, Russia, Singapore, Spain, Sweden, Switzerland, Thailand, Turkey, United Kingdom, & Vietnam.
Product codes and lots
P/N 9734724, 9734724K

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-3142-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-3139-2017Class IITerminatedVoluntary: Firm initiated
Z-3140-2017Class IITerminatedVoluntary: Firm initiated
Z-3141-2017Class IITerminatedVoluntary: Firm initiated
Z-3142-2017Class IITerminatedVoluntary: Firm initiated
Z-3143-2017Class IITerminatedVoluntary: Firm initiated