Z-3118-2017Class II

Sekisui Diagnostics Sekure Chemistry Acetaminophen L3K , Ref. 506-30. The firm names on the label are Sekisui...

Siemens Healthcare Diagnostics Inc. / initiated 2017-07-20 / Terminated / root cause: manufacturing process

Beckman Coulter Inc. began a recall of Sekisui Diagnostics Sekure Chemistry Acetaminophen L3K , Ref. 506-30. The firm names on the label are Sekisui Diagnostics (UK) Limited, United Kingdom, and... on . FDA classified it as Class II and gives the reason as "Reagent carryover from Sekisui ACTH when utilized on the BEC AU 680 analyzer may lead to falsely elevated BEC Lipase recovery exceeding the total precision specification". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-3118-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Sekisui Diagnostics Sekure Chemistry Acetaminophen L3K , Ref. 506-30. The firm names on the label are Sekisui Diagnostics (UK) Limited, United Kingdom, and Sekisui Diagnostics P.E.I. Inc., Canada. Used in laboratory testing.
Product code
LDP Colorimetry, Acetaminophen
Reason for recall
Reagent carryover from Sekisui ACTH when utilized on the BEC AU 680 analyzer may lead to falsely elevated BEC Lipase recovery exceeding the total precision specification.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The recalling firm issued letters dated 7/17/2017 via regular mail and email on 7/20/2017 to all customers receiving the affected product.
Quantity in commerce
320 units
Distribution
Distribution US nationwide.
Product codes and lots
All lots

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-3118-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-3118-2017Class IITerminatedVoluntary: Firm initiated