MODIFIED/SPECIAL INSTRUMENT, TURON GLENOID REAMER, REAM AND RUN, Catalog Numbers: REF S-200956 - 5mm PIN...
Encore Medical, Lp began a recall of MODIFIED/SPECIAL INSTRUMENT, TURON GLENOID REAMER, REAM AND RUN, Catalog Numbers: REF S-200956 - 5mm PIN, SIZE 46 REF S-200957 - 5mm PIN, SIZE 50 REF S-200958... on . FDA classified it as Class II and gives the reason as "It was reported that the models used to manufacture S-200956/958 had the wrong dimension used to create the spherical radius cutting surface. Device design allows a larger spherical radius cutting surface than intended". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-3102-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Encore Medical, LP
- Product
- MODIFIED/SPECIAL INSTRUMENT, TURON GLENOID REAMER, REAM AND RUN, Catalog Numbers: REF S-200956 - 5mm PIN, SIZE 46 REF S-200957 - 5mm PIN, SIZE 50 REF S-200958 - 5mm PIN, SIZE 58
- Product code
- HTO Reamer
- Reason for recall
- It was reported that the models used to manufacture S-200956/958 had the wrong dimension used to create the spherical radius cutting surface. Device design allows a larger spherical radius cutting surface than intended.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- The firm initiated their recall by telephone and email to their sales representatives and surgeons on 08/04/2017. The letter explained the problem and advised the surgeons that the sales representative would be collecting the devices for return.
- Quantity in commerce
- 5 units
- Distribution
- OH, VA
- Product codes and lots
- Catalog number, Lot Number S-200956, 535U1000 S-200957, 536U1000 S-200958, 537U1000
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-3102-2017 | Class II | Terminated | Voluntary: Firm initiated |