510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-3102-2017Class II

MODIFIED/SPECIAL INSTRUMENT, TURON GLENOID REAMER, REAM AND RUN, Catalog Numbers: REF S-200956 - 5mm PIN...

Encore Medical, LP / initiated 2017-08-04 / Terminated / root cause: manufacturing process

Encore Medical, Lp began a recall of MODIFIED/SPECIAL INSTRUMENT, TURON GLENOID REAMER, REAM AND RUN, Catalog Numbers: REF S-200956 - 5mm PIN, SIZE 46 REF S-200957 - 5mm PIN, SIZE 50 REF S-200958... on . FDA classified it as Class II and gives the reason as "It was reported that the models used to manufacture S-200956/958 had the wrong dimension used to create the spherical radius cutting surface. Device design allows a larger spherical radius cutting surface than intended". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-3102-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Encore Medical, LP
Product
MODIFIED/SPECIAL INSTRUMENT, TURON GLENOID REAMER, REAM AND RUN, Catalog Numbers: REF S-200956 - 5mm PIN, SIZE 46 REF S-200957 - 5mm PIN, SIZE 50 REF S-200958 - 5mm PIN, SIZE 58
Product code
HTO Reamer
Reason for recall
It was reported that the models used to manufacture S-200956/958 had the wrong dimension used to create the spherical radius cutting surface. Device design allows a larger spherical radius cutting surface than intended.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
The firm initiated their recall by telephone and email to their sales representatives and surgeons on 08/04/2017. The letter explained the problem and advised the surgeons that the sales representative would be collecting the devices for return.
Quantity in commerce
5 units
Distribution
OH, VA
Product codes and lots
Catalog number, Lot Number S-200956, 535U1000 S-200957, 536U1000 S-200958, 537U1000

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-3102-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-3102-2017Class IITerminatedVoluntary: Firm initiated