MyCareLink Smart Patient Monitors. It is intended for use with a compatible Medtronic patient implanted heart...
Medtronic Inc., Cardiac Rhythm and Heart Failure began a recall of MyCareLink Smart Patient Monitors. It is intended for use with a compatible Medtronic patient implanted heart device. The reader is a portable electronic... on . FDA classified it as Class II and gives the reason as "Patients monitored on two (2) or more implanted Medtronic heart devices in the Medtronic CareLink Network may have potential impact on the ability to remotely monitor the patient's heart devices. Potential impacts could...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0400-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- MyCareLink Smart Patient Monitors. It is intended for use with a compatible Medtronic patient implanted heart device. The reader is a portable electronic device that communicates with an implanted heart device
- Product code
- NVZ Pulse Generator, Permanent, Implantable
- Reason for recall
- Patients monitored on two (2) or more implanted Medtronic heart devices in the Medtronic CareLink Network may have potential impact on the ability to remotely monitor the patient's heart devices. Potential impacts could lead to missed CareAlert notifications or device reports.
- Root cause tag
- labeling and UDI
- FDA root cause
- Error in labeling
- Action
- Customers were notified via letter on 07/14/2017. Medtronic communicated that an updated Clinician Instructions for Use has been released with the following updated information: In patients with more than one Medtronic implanted hear device, device data may not be able to transmit from any of the implanted devices to the physician via the Monitor; The patient could potentially transmit device data unintentionally from one of the implanted heart devices when attempting to transmit from another device implanted in the same patient; and a Medtronic Programmer can be used to obtain data from each implanted heart device. Instructions included to ensure all appropriate personnel are notified of this correction.
- Quantity in commerce
- 1,519,984 enrollments total
- Distribution
- Domestic: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, Internationally: Austria, Belgium, Brunei, China, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Japan, Kazakhstan, Latvia, Lithuania, Malaysia, Malta, Netherlands, Norway, Philippines, Poland, Portugal, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, UAE, United Kingdom
- Product codes and lots
- Model 25000, UPN/GTIN 00643169527546, 00643169569744, 00643169666337, 00643169697843¿, 00643169862876, 00643169697850, 00643169742246, 00643169862883, 00643169666375, 00643169666382, 00643169666344, 00643169666351, 00643169666368, 00643169666320, 00643169590397, 00643169527584, 00643169527591, 00643169527553, 00643169527560, 00643169527577, 00643169947108
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0399-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0400-2018 | Class II | Terminated | Voluntary: Firm initiated |