Z-0400-2018Class II

MyCareLink Smart Patient Monitors. It is intended for use with a compatible Medtronic patient implanted heart...

Medtronic, Inc. / initiated 2017-07-14 / Terminated / root cause: labeling and UDI

Medtronic Inc., Cardiac Rhythm and Heart Failure began a recall of MyCareLink Smart Patient Monitors. It is intended for use with a compatible Medtronic patient implanted heart device. The reader is a portable electronic... on . FDA classified it as Class II and gives the reason as "Patients monitored on two (2) or more implanted Medtronic heart devices in the Medtronic CareLink Network may have potential impact on the ability to remotely monitor the patient's heart devices. Potential impacts could...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0400-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
MyCareLink Smart Patient Monitors. It is intended for use with a compatible Medtronic patient implanted heart device. The reader is a portable electronic device that communicates with an implanted heart device
Product code
NVZ Pulse Generator, Permanent, Implantable
Reason for recall
Patients monitored on two (2) or more implanted Medtronic heart devices in the Medtronic CareLink Network may have potential impact on the ability to remotely monitor the patient's heart devices. Potential impacts could lead to missed CareAlert notifications or device reports.
Root cause tag
labeling and UDI
FDA root cause
Error in labeling
Action
Customers were notified via letter on 07/14/2017. Medtronic communicated that an updated Clinician Instructions for Use has been released with the following updated information: In patients with more than one Medtronic implanted hear device, device data may not be able to transmit from any of the implanted devices to the physician via the Monitor; The patient could potentially transmit device data unintentionally from one of the implanted heart devices when attempting to transmit from another device implanted in the same patient; and a Medtronic Programmer can be used to obtain data from each implanted heart device. Instructions included to ensure all appropriate personnel are notified of this correction.
Quantity in commerce
1,519,984 enrollments total
Distribution
Domestic: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, Internationally: Austria, Belgium, Brunei, China, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Japan, Kazakhstan, Latvia, Lithuania, Malaysia, Malta, Netherlands, Norway, Philippines, Poland, Portugal, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, UAE, United Kingdom
Product codes and lots
Model 25000, UPN/GTIN 00643169527546, 00643169569744, 00643169666337, 00643169697843¿, 00643169862876, 00643169697850, 00643169742246, 00643169862883, 00643169666375, 00643169666382, 00643169666344, 00643169666351, 00643169666368, 00643169666320, 00643169590397, 00643169527584, 00643169527591, 00643169527553, 00643169527560, 00643169527577, 00643169947108

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0400-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0399-2018Class IITerminatedVoluntary: Firm initiated
Z-0400-2018Class IITerminatedVoluntary: Firm initiated