Z-0025-2018Class II

VCLAS (Visualase Cooled Laser Ablation System) Part Number: 9735559 Visualase Cooled Laser Applicator System...

Medtronic Navigation, Inc / initiated 2017-07-27 / Terminated / root cause: manufacturing process

Medtronic Navigation, Inc. began a recall of VCLAS (Visualase Cooled Laser Ablation System) Part Number: 9735559 Visualase Cooled Laser Applicator System 3mm Tip; Part Number: 9735560 Visualase Cooled... on . FDA classified it as Class II and gives the reason as "The VCLAS 3mm, 10mm, and 15mm tip devices were not appropriately tested for saline leakage during the manufacturing process". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0025-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Navigation, Inc
Product
VCLAS (Visualase Cooled Laser Ablation System) Part Number: 9735559 Visualase Cooled Laser Applicator System 3mm Tip; Part Number: 9735560 Visualase Cooled Laser Applicator System 10mm Tip; Part Number: 9735561 Visualase Cooled Laser Applicator System 15mm Tip
Product code
GEX Powered Laser Surgical Instrument
Reason for recall
The VCLAS 3mm, 10mm, and 15mm tip devices were not appropriately tested for saline leakage during the manufacturing process.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
All US consignees will be a sent a customer letter, dated July 27, 2017, via FedEx or Email. In addition, customers will be asked to return a signed consignee response form.
Quantity in commerce
423 devices
Distribution
US nationwide distribution.
Product codes and lots
Part Number: 9735559 Visualase Cooled Laser Applicator System 3mm Tip Lots - 000212782059, 000212823919, 000212894256, 000212978141, 000212783784, 000212854585, 000212926008, 000212996201, 000212823883, 000212882619, 000212956163, 000213145362; Part Number: 9735560 Visualase Cooled Laser Applicator System 10mm Tip Lots - 000212354781, 000212656015, 000212760821, 000212913269, 000212384009, 000212660280, 000212827856, 000212922387, 000212465594, 000212664497, 000212838837, 000212935278, 000212507737, 000212698642, 000212846727, 000213000097, 000212510629, 000212715926, 000212856596, 000213001255, 000212624409, 000212745633, 000212873863, 000213009483, 000212626439, 000212753561, 000212873873, 000213041517, 000212632885, 000212756095, 000212879480, 000213056030; Part Number: 9735561 Visualase Cooled Laser Applicator System 15mm Tip Lots - 000212141557, 000212503634, 000212647830, 000212796559, 000212148368, 000212504463, 000212701679, 000212846726, 000212434500, 000212549456, 000212707035, 000212901580, 000212438405, 000212553428, 000212745660, 000212912501, 000212473141, 000212632884, 000212795235, 000212963182