VCLAS (Visualase Cooled Laser Ablation System) Part Number: 9735559 Visualase Cooled Laser Applicator System...
Medtronic Navigation, Inc. began a recall of VCLAS (Visualase Cooled Laser Ablation System) Part Number: 9735559 Visualase Cooled Laser Applicator System 3mm Tip; Part Number: 9735560 Visualase Cooled... on . FDA classified it as Class II and gives the reason as "The VCLAS 3mm, 10mm, and 15mm tip devices were not appropriately tested for saline leakage during the manufacturing process". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0025-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic Navigation, Inc
- Product
- VCLAS (Visualase Cooled Laser Ablation System) Part Number: 9735559 Visualase Cooled Laser Applicator System 3mm Tip; Part Number: 9735560 Visualase Cooled Laser Applicator System 10mm Tip; Part Number: 9735561 Visualase Cooled Laser Applicator System 15mm Tip
- Product code
- GEX Powered Laser Surgical Instrument
- Reason for recall
- The VCLAS 3mm, 10mm, and 15mm tip devices were not appropriately tested for saline leakage during the manufacturing process.
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- All US consignees will be a sent a customer letter, dated July 27, 2017, via FedEx or Email. In addition, customers will be asked to return a signed consignee response form.
- Quantity in commerce
- 423 devices
- Distribution
- US nationwide distribution.
- Product codes and lots
- Part Number: 9735559 Visualase Cooled Laser Applicator System 3mm Tip Lots - 000212782059, 000212823919, 000212894256, 000212978141, 000212783784, 000212854585, 000212926008, 000212996201, 000212823883, 000212882619, 000212956163, 000213145362; Part Number: 9735560 Visualase Cooled Laser Applicator System 10mm Tip Lots - 000212354781, 000212656015, 000212760821, 000212913269, 000212384009, 000212660280, 000212827856, 000212922387, 000212465594, 000212664497, 000212838837, 000212935278, 000212507737, 000212698642, 000212846727, 000213000097, 000212510629, 000212715926, 000212856596, 000213001255, 000212624409, 000212745633, 000212873863, 000213009483, 000212626439, 000212753561, 000212873873, 000213041517, 000212632885, 000212756095, 000212879480, 000213056030; Part Number: 9735561 Visualase Cooled Laser Applicator System 15mm Tip Lots - 000212141557, 000212503634, 000212647830, 000212796559, 000212148368, 000212504463, 000212701679, 000212846726, 000212434500, 000212549456, 000212707035, 000212901580, 000212438405, 000212553428, 000212745660, 000212912501, 000212473141, 000212632884, 000212795235, 000212963182
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0025-2018 | Class II | Terminated | Voluntary: Firm initiated |