Power Processor 1K Stockyard. The Power Processor performs all pre-analytical sample tube preparation
Beckman Coulter Inc. began a recall of Power Processor 1K Stockyard. The Power Processor performs all pre-analytical sample tube preparation on . FDA classified it as Class II and gives the reason as "Beckman Coulter has identified that due to a PLC software nonconformity the 1K Stockyard can initiate a retrieval of a sample tube during the rack loading process, which should not occur. This issue and associated...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0046-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Power Processor 1K Stockyard. The Power Processor performs all pre-analytical sample tube preparation.
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- Beckman Coulter has identified that due to a PLC software nonconformity the 1K Stockyard can initiate a retrieval of a sample tube during the rack loading process, which should not occur. This issue and associated complaint were discovered and filed internally.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- BEC will initiate a SW update- to fix the SW defect. This will be executed via MOD (field modification) performed by Field Service.
- Quantity in commerce
- 27 units
- Distribution
- China France Italy Kuwait Spain Turkey United States
- Product codes and lots
- Software version PLC V26R10
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0046-2018 | Class II | Terminated | Voluntary: Firm initiated |