Z-0046-2018Class II

Power Processor 1K Stockyard. The Power Processor performs all pre-analytical sample tube preparation

Siemens Healthcare Diagnostics Inc. / initiated 2017-08-03 / Terminated / root cause: software

Beckman Coulter Inc. began a recall of Power Processor 1K Stockyard. The Power Processor performs all pre-analytical sample tube preparation on . FDA classified it as Class II and gives the reason as "Beckman Coulter has identified that due to a PLC software nonconformity the 1K Stockyard can initiate a retrieval of a sample tube during the rack loading process, which should not occur. This issue and associated...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0046-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Power Processor 1K Stockyard. The Power Processor performs all pre-analytical sample tube preparation.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Beckman Coulter has identified that due to a PLC software nonconformity the 1K Stockyard can initiate a retrieval of a sample tube during the rack loading process, which should not occur. This issue and associated complaint were discovered and filed internally.
Root cause tag
software
FDA root cause
Software design
Action
BEC will initiate a SW update- to fix the SW defect. This will be executed via MOD (field modification) performed by Field Service.
Quantity in commerce
27 units
Distribution
China France Italy Kuwait Spain Turkey United States
Product codes and lots
Software version PLC V26R10

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0046-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0046-2018Class IITerminatedVoluntary: Firm initiated