Z-3119-2017Class II
eMAG System, Ref 418591 It is an In Vitro Diagnostic Medical device intended for the automated isolation...
bioMerieux, Inc. / initiated 2017-04-10 / Terminated / root cause: manufacturing process
bioMerieux, Inc. began a recall of eMAG System, Ref 418591 It is an In Vitro Diagnostic Medical device intended for the automated isolation (purification and concentration) of total nucleic... on . FDA classified it as Class II and gives the reason as "Some anomalies have been identified during manufacturing controls". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-3119-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- bioMerieux, Inc.
- Product
- eMAG System, Ref 418591 It is an In Vitro Diagnostic Medical device intended for the automated isolation (purification and concentration) of total nucleic acids (RNA/DNA) from biological specimens, with liquid and homogeneous properties (as part of their natural characteristics or after pre-treatment).
- Product code
- JJH Clinical Sample Concentrator
- Reason for recall
- Some anomalies have been identified during manufacturing controls.
- Root cause tag
- manufacturing process
- FDA root cause
- Unknown/Undetermined by firm
- Action
- The firm, Biomerieux, sent an "IMPORTANT: URGENT PRODUCT SAFETY CORRECTION NOTICE" letter on 4/10/2017 to its Consignees. The letter described the product, problem and actions to be taken. The Consignees were instructed to distribute this information to all appropriate personnel; retain copy for files; forward information to all parties that may use this product; perform a Main Board Re-initialization after each weekly maintenance; and complete and return the Acknowledgment Form by Fax to + 33 (0) 04 78 87 73 07 to confirm receipt of notice. A Field Service engineer will come on site to check the unit. If you require additional assistance or have any questions, please contact your local bioMrieux Customer Service representative or call 919-620-3396 or email: [email protected].
- Quantity in commerce
- 15 US
- Distribution
- Worldwide Distribution: US (nationwide) to states of: GA, IL, IN, MA, NC & TX and to countries of: Austria, France, Germany, Hong Kong, Italy, Netherlands, Spain, Sweden, Switzerland, Thailand and United Kingdom.
- Product codes and lots
- Serial Numbers - IM03001, IM03002, IM3004, IM03005, IM03006, IM03007, IM03008, IM03009, IM03011, IM03012, IM03013, IM03014, IM03015, IM03017, IM03019
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-3119-2017 | Class II | Terminated | Voluntary: Firm initiated |