AQUIOS CL Flow Cytometer System The AQUIOS CL Flow Cytometer is intended for use as an in vitro diagnostic...
Beckman Coulter Inc. began a recall of AQUIOS CL Flow Cytometer System The AQUIOS CL Flow Cytometer is intended for use as an in vitro diagnostic device for immunophenotyping on . FDA classified it as Class II and gives the reason as "Beckman Coulter has identified that an incorrect setting using a proprietary software configuration tool can result in the Sample Prep Probe of the AQUIOS CL drawing aspirant from the wrong tube in a cassette which...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-3135-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- AQUIOS CL Flow Cytometer System The AQUIOS CL Flow Cytometer is intended for use as an in vitro diagnostic device for immunophenotyping.
- Product code
- OYE Flow Cytometric Reagents And Accessories.
- Reason for recall
- Beckman Coulter has identified that an incorrect setting using a proprietary software configuration tool can result in the Sample Prep Probe of the AQUIOS CL drawing aspirant from the wrong tube in a cassette which manifests in incorrect values being reported out of the laboratory for some patient samples.
- Root cause tag
- software
- FDA root cause
- Under Investigation by firm
- Action
- The firm, Beckman Coulter, sent an "URGENT MEDICAL DEVICE RECALL" letter dated August 16, 2017 to the affected customers via email, mail, and phone during the week of August 21, 2017. The letter described the product, problem and actions to be taken. The customers are instructed to follow instructions provided in the letter and complete and return the Customer Response Form within 10 days via Fax to: (786)-639-7500 /7501 /7502 /7504; mail to: Beckman Coulter, Inc.,- Attn: Regulatory Affairs Mail Code 31-B06; 11800 S.W. 147th Avenue; P O BOX 169015, Miami, FL 33116-9015 or email (if received electronically) at: [email protected] -include in subject field: FA-31696 AQUIOS CL August 2017. If you have any questions regarding this notification, please contact: From our website: http://www.beckmancoulter.com By phone: call 800-369-0333 in the United States and Canada. By email: [email protected] Outside the United States and Canada, contact your local Beckman Coulter representative.
- Quantity in commerce
- 213 units (34 units in the US)
- Distribution
- Worldwide Distribution: United States (nationwide) and countries of: Angola, Australia, Botswana, Canada, Chile, Columbia, Ghana, Israel, Malaysia, Namibia, Pakistan, Panama, South Africa, Uganda, France, Switzerland, Poland, Portugal, Germany, Belgium, Finland, Czech Republic, Denmark, Netherland, Turkey, Spain, Great Britain, Romania, Brazil, Italy, Sweden, Ghana, Colombia, Lesotho and Zambia
- Product codes and lots
- All software versions.; Serial numbers - BA30048, AY20032, BA01007, AZ19046, AZ12030, BA25046, BA05011, AY51072, BA01002, AZ15044, BA30049, BA30050, BA05015, AZ19047, BA18029, AY47071, AZ15038, BA23036, BA23035, AZ23058, AZ20057, AY47063, BA15024, BA15023, AZ45080, BA01010, AZ25070, AZ32071, AZ12033, AZ25065, AZ25068, AY51075, AZ19052, BA15025
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-3135-2017 | Class II | Terminated | Voluntary: Firm initiated |