AQUIOS CL Flow Cytometer System The AQUIOS CL Flow Cytometer is intended for use as an in vitro diagnostic...

Siemens Healthcare Diagnostics Inc. / initiated 2017-08-21 / Terminated / root cause: software

Beckman Coulter Inc. began a recall of AQUIOS CL Flow Cytometer System The AQUIOS CL Flow Cytometer is intended for use as an in vitro diagnostic device for immunophenotyping on . FDA classified it as Class II and gives the reason as "Beckman Coulter has identified that an incorrect setting using a proprietary software configuration tool can result in the Sample Prep Probe of the AQUIOS CL drawing aspirant from the wrong tube in a cassette which...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-3135-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
AQUIOS CL Flow Cytometer System The AQUIOS CL Flow Cytometer is intended for use as an in vitro diagnostic device for immunophenotyping.
Product code
OYE Flow Cytometric Reagents And Accessories.
Reason for recall
Beckman Coulter has identified that an incorrect setting using a proprietary software configuration tool can result in the Sample Prep Probe of the AQUIOS CL drawing aspirant from the wrong tube in a cassette which manifests in incorrect values being reported out of the laboratory for some patient samples.
Root cause tag
software
FDA root cause
Under Investigation by firm
Action
The firm, Beckman Coulter, sent an "URGENT MEDICAL DEVICE RECALL" letter dated August 16, 2017 to the affected customers via email, mail, and phone during the week of August 21, 2017. The letter described the product, problem and actions to be taken. The customers are instructed to follow instructions provided in the letter and complete and return the Customer Response Form within 10 days via Fax to: (786)-639-7500 /7501 /7502 /7504; mail to: Beckman Coulter, Inc.,- Attn: Regulatory Affairs Mail Code 31-B06; 11800 S.W. 147th Avenue; P O BOX 169015, Miami, FL 33116-9015 or email (if received electronically) at: [email protected] -include in subject field: FA-31696 AQUIOS CL August 2017. If you have any questions regarding this notification, please contact: From our website: http://www.beckmancoulter.com By phone: call 800-369-0333 in the United States and Canada. By email: [email protected] Outside the United States and Canada, contact your local Beckman Coulter representative.
Quantity in commerce
213 units (34 units in the US)
Distribution
Worldwide Distribution: United States (nationwide) and countries of: Angola, Australia, Botswana, Canada, Chile, Columbia, Ghana, Israel, Malaysia, Namibia, Pakistan, Panama, South Africa, Uganda, France, Switzerland, Poland, Portugal, Germany, Belgium, Finland, Czech Republic, Denmark, Netherland, Turkey, Spain, Great Britain, Romania, Brazil, Italy, Sweden, Ghana, Colombia, Lesotho and Zambia
Product codes and lots
All software versions.; Serial numbers - BA30048, AY20032, BA01007, AZ19046, AZ12030, BA25046, BA05011, AY51072, BA01002, AZ15044, BA30049, BA30050, BA05015, AZ19047, BA18029, AY47071, AZ15038, BA23036, BA23035, AZ23058, AZ20057, AY47063, BA15024, BA15023, AZ45080, BA01010, AZ25070, AZ32071, AZ12033, AZ25065, AZ25068, AY51075, AZ19052, BA15025

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-3135-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-3135-2017Class IITerminatedVoluntary: Firm initiated