Bard Magic 3 Antibacterial Hydrophilic Intermittent Catheter Female, 14 FR (30/box), Rx Only intended for...
C.R. Bard, Inc. began a recall of Bard Magic 3 Antibacterial Hydrophilic Intermittent Catheter Female, 14 FR (30/box), Rx Only intended for urinary bladder drainage in patients requiring... on . FDA classified it as Class II and gives the reason as "Specific lots labeled as an Antibacterial Hydrophilic Intermittent Catheter may lack the hydrophilic coating". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-3226-2017
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- C.R. Bard Inc
- Product
- Bard Magic 3 Antibacterial Hydrophilic Intermittent Catheter Female, 14 FR (30/box), Rx Only intended for urinary bladder drainage in patients requiring catheterization for management of incontinence, voiding dysfunction, and surgical procedures.
- Product code
- MJC Catheter, Urological (Antimicrobial) And Accessories
- Reason for recall
- Specific lots labeled as an Antibacterial Hydrophilic Intermittent Catheter may lack the hydrophilic coating.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- Customers were notified via letter on 07/28/2017. Instructions included to examine their inventory and identify affected product and notify customers if further distributed. Bard Medical Division asks that customers contact customer service at 1-800-526-4455 if any affected units do not meet their needs.
- Quantity in commerce
- 45,270 units
- Distribution
- Distribution to AL, AZ, CA, FL, GA, IL, IN, KS, KY, MA, MO, NC, NJ, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA.
- Product codes and lots
- Catalog No. 51514; Lot No. 53623024, 53623054, 53623047, 53623122; Exp Date 01/31/2020
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-3226-2017 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.