Z-0395-2018Class II

Measurement Specialties Reusable Temperature Probe Autoclavable. Sold as Esophageal/Rectal (Adult)...

Measurement Specialties, Inc. / initiated 2017-03-13 / Open, Classified / root cause: sterility

Measurement Specialties Inc began a recall of Measurement Specialties Reusable Temperature Probe Autoclavable. Sold as Esophageal/Rectal (Adult), Esophageal/Rectal (Pediatric), and Skin (Adult) Intended... on . FDA classified it as Class II and gives the reason as "The reusable temperature probes are provided with instructions for sterilization that have not been proven through a validation study". As of , the recall status on record is Open, Classified, and MDM tags the root cause as sterility.

Recall record

Recall number
Z-0395-2018
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Measurement Specialties, Inc.
Product
Measurement Specialties Reusable Temperature Probe Autoclavable. Sold as Esophageal/Rectal (Adult), Esophageal/Rectal (Pediatric), and Skin (Adult) Intended for use by a licensed physician to continuously monitor human esophageal, rectal, or skin temperature. They are designed to be used with any patient monitoring system designed to accept 400 Series temperature probes.
Product code
FLL Continuous Measurement Thermometer
Reason for recall
The reusable temperature probes are provided with instructions for sterilization that have not been proven through a validation study.
Root cause tag
sterility
FDA root cause
Other
Action
Customers were notified via mail or email on March 13, 2017. Additional customers were later identified and notified on June 14, 2017. Instructions included to disregard the sterilization instructions listed in the IFU (A08411F)originally provided with probes and instead utilize the updated IFU 08411G which contains validated sterilization instructions. They were also instructed to forward the recall notification to within their firm and to all organizations if further distributed. For further questions, please call (763) 316-4075.
Quantity in commerce
6,076 devices total
Distribution
USA (nationwide) distribution to the states of : AL, CA, KY, IL, KS, MN, NY, OH, TX, VA and WI. Distributed internationally to: Australia, Canada, China, England, Germany, India, Ireland, Israel, Italy, Japan, Malaysia, Netherlands, Singapore, South Africa, South Korea, Taiwan, Vietnam
Product codes and lots
Part No. 401AC (Lot No. 14K87246, 15M96863, 13E72645, 12D61386, 13D72302, 16B97593, 15C91286, 13B64377, 14J86547, 12B59416, 13A69334, 15K95972, 15A97196, 15E92377, 15F92378, 13L77336, 12D61023, 14G84704, 14K87499, 14F84193, 13L77335, 16B97594, 15C91419, 13B70167, 15C91448, 15E92375, 15K96046, 16A97197, 12B59529, 16B97798, 15K96125, 16B97595, 15L96777, 12D61829, 13D72644, 15A89998, 13F73449, 12M64134, 15L96745, 15D90215, 15H95035, 13F74403, 13L78489, 13L77336, 12H65961, 14G84878, 12F62179, 14J87014, 12B55182, 14F84444, 14J87081, 13L77336, 13L78489, 14F84400, 12J66632, 16A97380, 12M68130, 15E92376, 14L88737, 14L87904, 12K66633, 13H74994, 13H74995, 12B60056, 12C57285, 14H86266, 16A97592, 14J87164, 12D59168, 14M89765, 12A54328, 15A89854, 14F84634, 14G84624, 15M96829, 15K95896);    Part No. 402AC (Lot No. 13E73262, 15E92379, 12E57286, 12L68344, 15E93385, 12H65752, 12H65787, 15F92380, 12J65322, 15F92381, 12K67454, 15A89941, 15A89282, 13H75930, 12F55548, 12K67533, 13H75930, 12J66809, 12B59570, 15E92010, 12A50150, 14E83393);    Part No. 409AC (Lot No. 12C56215, 12B56215, 15C90920, 15K28358, 12C60069, 13B70035, 16B97312, 13A69520, 14E83602, 13E73264, 12L68584, 11M57610, 12F59571, 15B90832, 15E92560, 12K63370, 16B97614, 12L54648, 15D92011, 12M63372, 14H85773, 16B97615, 15M97109, 16B97805, 12M69229, 15A89939, 15A90196, 12F60692, 14L88520, 14L88519, 13A70145, 15E92382, 15A90466, 13E73450, 15F92383, 13D72925)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0395-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0395-2018Class IIOngoingVoluntary: Firm initiated